The FDA announced on Oct. 13, 2026, that Tandem Diabetes Care is recalling the Tandem Mobi Insulin Pump with Interoperable Technology due to a software defect in the Android app. The hazard involves incorrect insulin delivery amounts. Owners should contact the company for details.
How to tell if you have it
- Brand: Tandem Diabetes Care
- Product: Tandem Mobi Insulin Pump with Interoperable Technology (with Android Tandem Mobi Mobile app)
- Model/Catalog Number: 1015472
- Lot Code: UDI-DI: 00389152547244
- Software versions: 3.5.4, 3.5.2.2, and 3.5.2.1
The hazard
A software defect in the Tandem Mobi mobile application for Android causes the immediate portion of an Extended Bolus to be incorrectly reduced to 0 units when the entered glucose value equals the user's active target glucose value. The deferred portion is delivered, but the total insulin delivered may be less than intended without a distinct alert. This potentially results in unrecognized under-delivery of insulin and hyperglycemia.
Context
GovInfoNow's records show Tandem Diabetes Care has had 12 other recalls since April 23, 2018. Recent recalls include the Sept. 23, 2026 t:slim X2 Insulin Pump recall for potential battery issues and the July 16, 2026 Tandem t:slim App recall regarding pairing issues with Abbott sensors. Other recent recalls of similar products include the May 27, 2026 Medtronic MiniMed 780G Insulin Pump recall for software errors causing insulin stoppage and the Feb. 13, 2026 Medtronic MiniMed 630G Insulin Pump recall for risk of insulin delivery errors.