Skip to content
Use caution Medical device recall · FDA · Published May 27, 2026

Medtronic MiniMed 780G Insulin Pump Recall: Software Error Causes Insulin Stoppage

Medtronic recalls MiniMed 780G insulin pumps with specific software versions due to a software anomaly that can cause insulin delivery to stop when paired with Instinct sensors, risking hyperglycemia or diabetic ketoacidosis.

🇺🇸 FDA recall #Z-2698-2026Medical deviceMedtronic ›
Take action

Contact the company

Recalling firmMedtronic MiniMed, Inc.18000 Devonshire St · Northridge, CA 91325-1219
✎ GovInfoNow plain-language explainer

Medtronic is recalling certain MiniMed 780G insulin pumps with software version 6.62. These pumps can stop delivering insulin when paired with Instinct sensors, potentially leading to high blood sugar levels or diabetic ketoacidosis if not addressed. Read more at govinfonow.com/recall/49068-medtronic-minimed-780g-insulin-pump-recall-software-error-causes-insulin.

Source Summary by GovInfoNow · govinfonow.com/recall/49068-medtronic-minimed-780g-insulin-pump-recall-software-error-causes-insulin · Updated August 10, 2026 · FDA enforcement record: FDA #Z-2698-2026

Cite this page

“Medtronic MiniMed 780G Insulin Pump Recall: Software Error Causes Insulin Stoppage.” GovInfoNow, updated August 10, 2026. https://govinfonow.com/recall/49068-medtronic-minimed-780g-insulin-pump-recall-software-error-causes-insulin. Based on FDA recall #Z-2698-2026 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2698-2026%22&limit=1).

The FDA announced on May 27, 2026, that Medtronic MiniMed, Inc. is recalling MiniMed 780G Insulin Pumps due to a software anomaly that may cause unexpected insulin delivery stops. The agency issued recall number Z-2698-2026 for the Class II voluntary recall affecting units distributed in Spain, Great Britain, the Netherlands, and Italy.

How to tell if you have it

  • MiniMed 780G Insulin Pump
  • REF: MMT-1885, MMT-1886, MMT-1895, MMT-1896
  • Software version 6.62
  • UDI-DI/GTIN: A7630006392701, 00763000521554, 00763000639273, A7630006392301, 00199150035687, A7630007325603, A7630007325601, 00763000639235, A7630007325604, 00763000384289, 00763000521530, A7630007325401, 00763000521547, A7630006392601, A7630006392201, 00763000639228, 00199150035663, 00763000639266, A7630007325404, 00199150035694, 00763000442194, A7630007325403, 00763000521523

The hazard

According to the FDA, a software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering insulin and displays two error messages back-to-back: Pump Error 53 followed immediately by Pump Error 23. Insulin delivery will stop which may lead to delayed therapy and an increased risk of hyperglycemia or diabetic ketoacidosis (DKA).

Context

GovInfoNow's records show Medtronic MiniMed, Inc. has 13 other recalls on record since November 15, 2023. Other recent recalls of similar products include Beta Bionics iLet Bionic Pancreas Software 1.4.2/1.4.3 Recall: Delayed Glucose Readings Risk (March 12, 2026), Insulet Omnipod DASH Pod Recall: Insulin Leaking Risk (May 20, 2026), and Tand: Pumps with Czech guide errors may cause insulin delivery errors (January 30, 2026). FDA's MAUDE device-report database lists 11,199 reports naming MiniMed 780G received since May 2025, including 2,900 marked serious and three listing a death.

How to tell if you have this one

Check for the product reference numbers MMT-1885, MMT-1886, MMT-1895, or MMT-1896 on the pump. The software version should be 6.62 if it's affected.

  • Brand
    Medtronic
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Pump Software Version

    Look for the software version on the pump's display; if it shows 6.62, it may be affected.

  2. Contact Medtronic Support

    Reach out to Medtronic for guidance on whether your pump needs a software update or replacement.

How this recall unfolded

Recall timeline

FDA published this recall 49 days after the firm started it.

1. Firm started the recall: May 27, 202612. FDA classified it Class II: Jul 8, 202623. Published in FDA’s Enforcement Report: Jul 15, 202632026
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Other Medtronic recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-2698-2026 (May 27, 2026) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-2698-2026 to see it in context.
  • Plain-language summary written by qwen3:4b on August 10, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA