The FDA announced on September 11, 2026, that Medtronic MiniMed is recalling CareLink Clinic software version 4.4A due to a defect causing incorrect insulin parameter calculations. Owners should contact the company for remedy details.
How to tell if you have it
- Brand: Medtronic MiniMed
- Product: CareLink Clinic Software
- Reference: MMT-7350
- Software Version: 4.4A
- UDI-DI: 0763000B00008748K
- Distribution: US Nationwide distribution in the state of KS
The hazard
According to the FDA, the software defect causes the Settings Calculator to ignore custom values entered by healthcare professionals for Total Daily Dose or Long-acting insulin. This results in incorrect rapid-acting insulin parameters that could lead to over- or under-delivery of insulin with severity ranging from catastrophic death to minor self-treatable conditions.
Context
GovInfoNow's records show Medtronic MiniMed, Inc. has 16 other recalls on record since November 15, 2023. Other recent recalls of similar products include Medtronic MiniMed Go App Recall: Missed Insulin Doses (September 2, 2026), MiniMed Flex Insulin Pump Recall: Inaccurate Delivery Risk (August 20, 2026), and Medtronic MiniMed 640G Insulin Pump Recall: Risk of Incorrect Insulin Delivery (July 13, 2026). Other recent recalls of similar products include Tand: Pumps with Czech guide errors may cause insulin delivery errors (January 30, 2026), Nova Biomedical StatStrip Glucose Meter Software Issue: Patient Data Mismatch Risk (July 29, 2026), and Medtronic SynchroMed Flex Infusion Software Issue: Steps May Not Program Correctly (April 8, 2026).