Skip to content
Use caution Medical device recall · FDA · Published November 19, 2019

Dexcom G6 CGM App iOS Software Version 1.4.0: Low Alarm Not Alerting After 30 Days

Dexcom recalls iOS software version 1.4.0 of its G6 CGM app up to 1.4.4 due to low alarm feature not alerting users after 30 days of app installation. This could lead to undetected high blood sugar levels.

🇺🇸 FDA recall #Z-1104-2020Medical deviceDexcom ›
Take action

Contact the company

Recalling firmDexcom Inc6340 Sequence Dr · San Diego, CA 92121-4356
✎ GovInfoNow plain-language explainer

What happened, in plain English

Dexcom is recalling iOS software versions 1.4.0 up to 1.4.4 for its G6 CGM app. The low alarm feature may not alert users if they enabled the Alert Schedule feature more than 30 days after installing the app, potentially leading to undetected high blood sugar levels. Read more at govinfonow.com/recall/66500-dexcom-g6-cgm-app-ios-software-version-1-4-0.

Source Summary by GovInfoNow · govinfonow.com/recall/66500-dexcom-g6-cgm-app-ios-software-version-1-4-0 · Updated August 19, 2026 · FDA enforcement record: FDA #Z-1104-2020

Cite this page

“Dexcom G6 CGM App iOS Software Version 1.4.0: Low Alarm Not Alerting After 30 Days.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/66500-dexcom-g6-cgm-app-ios-software-version-1-4-0. Based on FDA recall #Z-1104-2020 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1104-2020%22&limit=1).

Why it’s dangerous

The low alarm feature in the app may not alert users when their blood sugar levels are high if they enabled the Alert Schedule feature more than 30 days after installing the app. This could lead to undetected high blood sugar levels, which might cause health complications.

Who’s affected

People using the Dexcom G6 CGM app on iOS who enabled the Alert Schedule feature more than 30 days after installing the app are affected. Those with diabetes who rely on continuous glucose monitoring are most at risk.

How to tell if you have this one

Check if you have the Dexcom G6 CGM app on iOS with software version 1.4.0 up to 1.4.4. The product was sold with the GTIN 00386270000804.

  • Brand
    Dexcom
  • Product
    Dexcom Inc recalls Brand Name: Dexcom G6 CGM App for iOS Software Number: SW11677 Software Version: Version 1.4.0 up to corrected Version
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Update Software

    Update the app to version 1.4.4 or later to fix the low alarm issue.

Where it went and when

Where it was distributed

Distributed nationwide, and outside the U.S.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “US Nation-wide OUS: Austria; Belgium; Canada; Czechia; Denmark; Finland; Germany; Great Britain; Hong Kong; Ireland; Italy; Luxembourg; Netherlands; Norway; South Africa; Spain; Sweden, and Switzerland.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide, and outside the U.S.

Recall timeline

FDA terminated this recall 3.8 years after the firm started it.

1. Firm started the recall: Nov 19, 201912. FDA classified it Class II: Feb 7, 202023. Published in FDA’s Enforcement Report: Feb 19, 202034. FDA terminated the recall: Aug 22, 2023420192023
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

Similar recalls

Other Dexcom recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1104-2020 (November 19, 2019) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1104-2020 to see it in context.