The FDA announced on November 14, 2022, that Insulet Corporation is voluntarily recalling Omnipod 5 Automated Insulin Delivery Systems due to melting or deforming charging ports. The hazard includes potential minor burns and failure to charge the device.
How to tell if you have it
- Brand: Insulet
- Product: Omnipod 5 Automated Insulin Delivery System
- Catalog numbers: PT-000409, PT-000428, PT-000429
- Units affected: 129,086
The hazard
The Omnipod 5 Controller charging port and cable may melt, deform, or discolor from heat generated by increased resistance. This excess heat may cause minor burns if touched, and the controller may fail to charge, requiring users to switch to a backup insulin plan.
Context
GovInfoNow's records show Insulet Corporation has 11 other recalls on record since December 18, 2018. Other recent recalls of similar products include Insulet Omnipod 5 Pods Recall: Insulin Leak Risk from Manufacturing Defect (May 20, 2026), Insulet Omnipod 5 Pods Recall: Internal Leak Risk from Manufacturing Defect (March 12, 2026), and Insulet Omnipod 5 App: Insulin Entry Errors Risk Hypoglycemia (September 11, 2025).