The FDA announced on March 12, 2026, that Insulet Corporation is recalling Omnipod 5 Pods due to a manufacturing defect causing insulin leakage. Owners should contact the company for a replacement, according to the FDA.
How to tell if you have it
- Brand: Insulet
- Product: Omnipod 5 Pods
- Model/Catalog number: PT-001662 ASM 5PK Pod STRL OPS G6G7
- UDI-DI: 10385083000527
- Lot Codes: PH1U02252541, PH1U03282511, PH1U03282522, PH1U03312511, PH1U03312521, PH1U04012511, PH1U04012521, PH1U05052511, PH1U08162531, PH1U08182531, PH1U08182541, PH1U09242511, PH1U09242521, PH1U09242531, PH1U09252521, PH1U09252531, PH1U10152541, PH1U10162531, PH1U10162541, PH1U10172531, PH1U10172541, PH1U10182531, PH1U10182541, PH1U10202511, PH1U10202521, PH1U10202531, PH1U10202541, PH1U10212531, PH1U10212541, PH1U10222531, PH1U10222541, PH1U10232531, PH1U10232541, PH1U10242521, PH1U10242531, PH1U10242541
- Units affected: 1,240,115
The hazard
The defect involves an internal soft cannula tear that results in insulin leaking into the Pod instead of being delivered to the user. This can cause pump under-delivery or cessation of insulin delivery, putting users at risk of hyperglycemia, diabetic ketoacidosis (DKA), or death. Not all devices with the defect will issue an alarm.
Context
GovInfoNow's records show Insulet Corporation has 11 other recalls on record since 2018-12-18. Other recent recalls of similar products include Insulet Omnipod Eros Pod Recall (2026-05-20), Insulet Omnipod 5 App Recall (2025-09-11), and Insulet Omnipod 5 iOS App Recall (2025-05-23). FDA's MAUDE device-report database contains 106,945 reports naming Omnipod 5 received since March 2025, including 11,434 marked serious and 114 listing a death. Other recent recalls of similar products include Medtronic MiniMed 780G Insulin Pump Recall (2026-07-13), Medtronic MiniMed 640G Insulin Pump Recall (2026-07-13), and Medtronic MiniMed 720G Insulin Pump Recall (2026-02-13).