The FDA announced on July 3, 2025, that Tandem Diabetes Care, Inc. is recalling the t:slim X2 Insulin Pump with Interoperable Technology due to an app defect causing pairing issues and graphical errors. These defects can potentially lead to incorrect therapy decisions and subsequent hypoglycemia or hyperglycemia.
How to tell if you have it
- Brand: Tandem
- Product: t:slim X2 Insulin Pump with Interoperable Technology (with Tandem t:slim mobile app)
- Software Version: Android Software Version 2.9.1
- UDI-DI code: 00389152470641
- Serial Numbers: 1163923, 1164430, 1167131, 1299238, 1300180, 1329242, 1381313, 1381315, 1381537, 1381570, 1391808, 1391892, 1392863, 1392953, 1393040, 90615975, 90754587, 901144584
- Affected Devices: Google Pixel 8 Pro, samsung SM-S901E, samsung SM-A346E, samsung SM-G780G, samsung SM-S928B, Google Pixel 6 Pro, Google Pixel 9, samsung SM-A546E, Xiaomi 2407FPN8EG, samsung SM-S711B, samsung SM-N980F, samsung SM-S926B, samsung SM-S916B, realme RMX3241
- Quantity: 19 units
- Distribution: International distribution to the country of Israel
The hazard
An app defect occurs when the phone is set to a right-to-left language, causing app-pump pairing issues and graphical defects. These errors can potentially lead to incorrect therapy decisions and subsequent hypoglycemia or hyperglycemia.
Context
GovInfoNow's records show Tandem Diabetes Care, Inc. has 10 other recalls on record since April 23, 2018. Other recent recalls of similar products include a July 16, 2026 recall of the t:slim Mobile Application for iOS pairing issues and an Oct. 6, 2025 recall of the Tandem Mobi Insulin Pump for false vibration motor failure risk. Other recent recalls of similar products include a May 27, 2026 recall of the Medtronic MiniMed 780G Insulin Pump for software error causing insulin stoppage and Feb. 13, 2026 recalls of Medtronic Paradigm and MiniMed 630G Insulin Pumps for risks of hypoglycemia or delivery errors.