GE Healthcare recalls Carescape Central Station V3 MAS800 Desktop software version V3.0.5 with part number 5867474-03. This version can cause system reboots when 118-120 central stations are connected, risking patient monitoring loss. Affected units have specific serial numbers listed in the recall.
Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.
GE Healthcare is recalling Carescape Central Station V3 MAS800 Desktop units with software version V3.0.5 (part number 5867474-03) because the software can enter a continuous reboot cycle when 118-120 central stations are connected to the network. This may cause loss of central patient monitoring. The recall affects specific units with listed serial numbers.
The software version V3.0.5 can cause the system to enter a continuous reboot cycle when 118, 119, or 120 central stations are connected to the network. This may result in loss of central monitoring of connected patients, potentially delaying critical medical interventions.
Healthcare facilities using the Carescape Central Station V3 MAS800 Desktop with software version V3.0.5 (part number 5867474-03) are most at risk. The issue occurs only when 118-120 central stations are connected to the network, which is uncommon in typical installations.
Check if your Carescape Central Station V3 MAS800 Desktop has software version V3.0.5 and part number 5867474-03. Affected units have specific serial numbers listed in the recall notice, including STV25225456SA and other variants.
Verify your Carescape Central Station V3 MAS800 Desktop software version is V3.0.5 with part number 5867474-03.
Ensure your system is not connected to a network with 118-120 other central stations, as this triggers the reboot cycle.
The software version V3.0.5 can cause the system to enter a continuous reboot cycle when 118-120 central stations are connected to the network, risking loss of patient monitoring.
The recall includes specific units with serial numbers listed in the notice, including STV25225456SA and other variants.
If your system is connected to 118-120 other central stations, monitor for reboots and contact GE Healthcare for software updates.
We’ve drafted a message you can send GE Healthcare to request your refund or repair — edit it as you like.
Subject: Recall Z-0214-2026 — GE Healthcare GE Healthcare Hello, I own a GE Healthcare that is covered by recall Z-0214-2026 from GE Healthcare. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.