GE Healthcare recalls Carescape Central Station V3 MAS800 Desktop software version V3.0.5 with part number 5867474-04. This version can cause system reboots when 118-120 central stations are connected, risking patient monitoring loss.
Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.
GE Healthcare is recalling its Carescape Central Station V3 MAS800 Desktop with software version V3.0.5. This version can cause the system to enter a continuous reboot cycle when 118, 119, or 120 central stations are connected to the network, potentially leading to loss of patient monitoring.
The software version V3.0.5 can cause the system to repeatedly reboot when 118, 119, or 120 central stations are connected to the network. This may interrupt central monitoring of patients, potentially causing safety risks if the system fails to track patient status.
Healthcare facilities using the Carescape Central Station V3 MAS800 Desktop with software version V3.0.5 (part number 5867474-04) are most at risk when connecting 118-120 central stations to the network.
Check for the Carescape Central Station V3 MAS800 Desktop with software version V3.0.5 (part number 5867474-04). Serial numbers listed in the recall include STU25210418SA and others.
Verify the software version on your Carescape Central Station V3 MAS800 Desktop. If it is V3.0.5, it is affected.
Ensure your facility does not have 118, 119, or 120 central stations connected to the network when using this software version.
The recalled software version is V3.0.5 for the Carescape Central Station V3 MAS800 Desktop.
If the system enters a continuous reboot cycle, it could cause loss of central monitoring of connected patients, potentially leading to safety risks.
Ensure your facility does not have 118, 119, or 120 central stations connected to the network when using this software version.
We’ve drafted a message you can send GE Healthcare to request your refund or repair — edit it as you like.
Subject: Recall Z-0213-2026 — GE Healthcare GE Healthcare Hello, I own a GE Healthcare that is covered by recall Z-0213-2026 from GE Healthcare. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.