The FDA announced on September 11, 2025, that Insulet Corporation is recalling Omnipod 5 iOS application software versions 2.X.X due to a hazard of insulin overdelivery. The defect causes digit duplication when users magnify the screen while entering numbers, potentially leading to hypoglycemia.
How to tell if you have it
- Brand: Insulet
- Product: Omnipod 5 App Software
- Software versions: 2.X.X
- UDI: 10385083000459
The hazard
The hazard occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field to magnify the visual field using the iPhone's magnifying feature while simultaneously selecting a single number on the keyboard. This causes the single digit number to duplicate, resulting in errors in manual entry of digits such as bolus entry or I:C ratio. Therefore, this may result in over delivery of insulin which may cause hypoglycemia.
Context
GovInfoNow's records show Insulet Corporation has had 11 other recalls since December 18, 2018. Recent similar recalls include the Insulet Omnipod Eros Pod Recall (May 20, 2026) for insulin leaking risk and the Insulet Omnipod 5 Pods Recall (March 12, 2026) for internal leak risk. Other recent recalls of similar products include the Medtronic InPen App recall (November 13, 2025), the Medtronic MiniMed 780G Insulin Pump Recall (May 27, 2026), and the Medtronic MiniMed Go App Version 1.0.0/1.1.0 recall (June 16, 2026).