Angiodynamics Accu2i pMTA Applicator Recall: Microwave Energy Failure Risk

Angiodynamics recalls Accu2i pMTA Applicators with potential microwave energy failure due to coolant ingress causing error codes. Affected units include specific model numbers and lots.

🇺🇸 FDA recall #Z-0531-2018Medical deviceAngiodynamics ›
Take action

Contact the company

Recalling firmAngiodynamics, Inc.603 Queensbury Ave · Queensbury, NY 12804-7619
✎ GovInfoNow plain-language explainer

What happened, in plain English

Angiodynamics is recalling Accu2i pMTA Microwave Tissue Ablation Applicators that may not function properly during use. This happens because coolant enters an electrical connection, causing the device to display a 'High Reflective Power' error code instead of delivering the intended microwave energy. Read more at govinfonow.com/recall/73947-angiodynamics-accu2i-pmta-applicator-recall-microwave-energy-failure-risk.

Source Summary by GovInfoNow · govinfonow.com/recall/73947-angiodynamics-accu2i-pmta-applicator-recall-microwave-energy-failure-risk · Updated September 23, 2026 · FDA enforcement record: FDA #Z-0531-2018

Cite this page

“Angiodynamics Accu2i pMTA Applicator Recall: Microwave Energy Failure Risk.” GovInfoNow, updated September 23, 2026. https://govinfonow.com/recall/73947-angiodynamics-accu2i-pmta-applicator-recall-microwave-energy-failure-risk. Based on FDA recall #Z-0531-2018 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0531-2018%22&limit=1).

Why it’s dangerous

Coolant can enter the electrical connection, causing the device to display a 'High Reflective Power' error code. This means the applicator won't deliver the intended microwave energy during tissue ablation procedures.

Who’s affected

Medical professionals using the Accu2i pMTA Applicator with Sulis V Generators are affected. The risk is low as the issue only affects the device's functionality, not causing injury.

How to tell if you have this one

Check for model number H787900602US0 and catalog number 900-602-US. The product was sold with Sulis V Generator REF 806-10 and includes specific production lots listed in the recall notice.

  • Brand
    Angiodynamics
  • Category
    Medical device
FDA recall event · 6 products

All 6 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class II
  • Status
    Terminated
  • Distribution
    Worldwide distribution - US Nationwide
Recall numberProduct, codes & lots
Z-0526-2018Accu2i pMTA Applicator, 14 cm STANDARD, REF/Catalog # 900-600, Model # (UPN) H7879006000, STERILE --- This is only for use with a Sulis V Generator, REF 806-10. Product Usage: The Acculis Accu2i pMTA Microwave Tissue Ablation Applicator is indicated for the intraoperative coagulation of soft tissue.; Lots: 15310411, 15320416, 15340428, 15350430, 15360434, 15380439, 15400449, 15420453, 15430454, 15440459, 15460466, 15470468, 15490476, 15500480, 15520487, 16010488, 16050494, 16070499, 16080502, 16100508, 16120512, 16140517, 16150520, 16170528, 16180530, 16190535, 16210540, 16230549, 16240550, 16250556, 16260561, 16270562, 16300568, 16310572, 16320573, 16350581, 16370590, 16380592, 16390595, 16400599, 16410603, 16420604, 16440611, 16450617, 17010645, 17020649, 17030653, 17040657, 17060662, 17070666, 17090673, 17100675, 17130686*, 17130687.
Z-0527-2018Accu2i pMTA Applicator, 29 cm LONG, REF/Catalog # 900-601, Model # (UPN) H7879006010, STERILE --- This is only for use with a Sulis V Generator, REF 806-10. Product Usage: The Acculis Accu2i pMTA Microwave Tissue Ablation Applicator is indicated for the intraoperative coagulation of soft tissue.; Lots: 15310414, 15320417, 15340425, 15360433, 15370437, 15380441, 15390445, 15400451, 15470470, 15480472, 16050496, 16070501, 16130515, 16160525, 16190536, 16250558, 16320574, 16330579, 16420606, 16440612, 16460620, 16480630, 16500637, 17010646, 17030654, 17040659, 17070667, 17090672.
Z-0528-2018Accu2i pMTA Applicator, 19 cm INTERMEDIATE, REF/Catalog # 900-602, Model # (UPN) H7879006020, STERILE --- This is only for use with a Sulis V Generator, REF 806-10. Product Usage: The Acculis Accu2i pMTA Microwave Tissue Ablation Applicator is indicated for the intraoperative coagulation of soft tissue.; Lots: 15310413, 15320415, 15340429, 15350431, 15360432, 15370435, 15380440, 15390444, 15400450, 15430456, 15440460, 15460467, 15470469, 15480471, 15490477, 15500481, 15520484, 16010489, 16050495, 16070500, 16080503, 16090506, 16100509, 16120513, 16150521, 16160523, 16170529, 16190534, 16230548, 16240551, 16250557, 16270563, 16300569, 16310571, 16320575, 16330577, 16350582, 16350586, 16370591, 16390596, 16410602, 16430610, 16440613, 16450618, 16460619, 16500636, 16510644, 17020651, 17040658, 17050660, 17060663, 17070668, 17100674, 17110678, 17130685.
Z-0529-2018Accu2i pMTA Applicator, 14 cm STANDARD, REF/Catalog # 900-600-US, Model # (UPN) H787900600US0, STERILE --- This is only for use with a Sulis V Generator, REF 806-10. Product Usage: The Acculis Accu2i pMTA Microwave Tissue Ablation Applicator is indicated for the intraoperative coagulation of soft tissue.; Lots: 15310412, 15330419, 15340427, 15390446, 15430455, 15440461, 15450463, 15480473, 15490478, 16030490, 16060498, 16130514, 16160524, 16180531, 16240552, 16260560, 16330578, 16350585, 16380593, 16400600, 16420605, 16430609, 16440614, 16460621, 16500638, 16510642, 17010647, 17020650, 17030655, 17050661, 17060664, 17080669, 17110676.
Z-0530-2018Accu2i pMTA Applicator, 29 cm LONG, REF/Catalog # 900-601-US, Model # (UPN) H787900601US0, STERILE --- This is only for use with a Sulis V Generator, REF 806-10. Product Usage: The Acculis Accu2i pMTA Microwave Tissue Ablation Applicator is indicated for the intraoperative coagulation of soft tissue.; Lots: 15330421, 15340426, 15370438, 15390447, 15400452, 15430458, 15440462, 15450465, 15480475, 15500482, 15520486, 16060497, 16100507, 16130516, 16140519, 16160526, 16180532, 16190537, 16240553, 16330580, 16350583, 16380594, 16410601, 16420607, 16430608, 16440616, 16460623, 16480629, 16500639, 17020652, 17090671, 17140694.
Z-0531-2018Accu2i pMTA Applicator, 19 cm INTERMEDIATE, REF/Catalog # 900-602-US, Model # (UPN) H787900602US0, STERILE --- This is only for use with a Sulis V Generator, REF 806-10. Product Usage: The Acculis Accu2i pMTA Microwave Tissue Ablation Applicator is indicated for the intraoperative coagulation of soft tissue.; Lots: 15330420, 15370436, 15390448, 15430457, 15450464, 15480474, 15490479, 15520485, 16030491, 16140518, 16160527, 16180533, 16300570, 16320576, 16350584, 16440615, 16460622, 16480628, 16500640, 16510643, 17010648, 17030656.
Based on the official FDA remedy

What to do — step by step

  1. Check Device Status

    Look for the error code 'High Reflective Power' on the microwave generator display to confirm if the device is malfunctioning.

Where it went and when

Where it was distributed

Distributed nationwide, and outside the U.S.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “Worldwide distribution - US Nationwide” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide, and outside the U.S.

Recall timeline

FDA terminated this recall 15 months after the firm started it.

1. Firm started the recall: Jun 5, 201712. FDA classified it Class II: Feb 5, 201823. Published in FDA’s Enforcement Report: Feb 14, 201834. FDA terminated the recall: Sep 11, 2018420172018
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0531-2018 (June 5, 2017) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0531-2018 to see it in context.
  • Plain-language summary written by qwen3:4b on September 23, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA