Skip to content
Use caution Medical device recall · FDA · Published June 15, 2017

Fujifilm FDR Visionary Suite CH-200 X-ray Tube Risk: Cracks May Cause Injury

Fujifilm recalls FDR Visionary Suite X-ray systems with CH-200 model due to potential cracks in the tube holding shaft that could lead to injury. Affected units have specific serial numbers and UDI codes.

🇺🇸 FDA recall #Z-0232-2018Medical deviceFujifilm Medical Systems ›
Take action

Contact the company

Recalling firmFujifilm Medical Systems U.S.A., Inc.419 West Ave · Stamford, CT 06902-6343
✎ GovInfoNow plain-language explainer

What happened, in plain English

Fujifilm is recalling X-ray systems with the CH-200 model because cracks in the tube holding shaft near the base of the mounting flange could cause injury. This affects specific units with serial numbers MP95A9F6B001, MP95A9F6A001, and MP95A8F5A001. Read more at govinfonow.com/recall/73853-fujifilm-fdr-visionary-suite-ch-200-x-ray-tube-risk.

Source Summary by GovInfoNow · govinfonow.com/recall/73853-fujifilm-fdr-visionary-suite-ch-200-x-ray-tube-risk · Updated September 23, 2026 · FDA enforcement record: FDA #Z-0232-2018

Cite this page

“Fujifilm FDR Visionary Suite CH-200 X-ray Tube Risk: Cracks May Cause Injury.” GovInfoNow, updated September 23, 2026. https://govinfonow.com/recall/73853-fujifilm-fdr-visionary-suite-ch-200-x-ray-tube-risk. Based on FDA recall #Z-0232-2018 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0232-2018%22&limit=1).

Why it’s dangerous

Cracks in the tube holding shaft near the base of the mounting flange could lead to the X-ray tube assembly becoming unstable. This might cause injury to users or patients during operation if the crack worsens over time.

Who’s affected

Medical facilities using the FDR Visionary Suite X-ray systems with CH-200 model are most at risk. Patients and staff could be affected if the device is used without addressing the crack issue.

How to tell if you have this one

Check for serial numbers MP95A9F6B001, MP95A9F6A001, or MP95A8F5A001 on the device. The product is part of the FDR Visionary Suite X-ray system with model CH-200.

  • Brand
    Fujifilm Medical Systems
  • Product
    FUJIFILM FDR Visionary Suite with CH-200. Model # CH-200, Catalog: Visionary Suite (FMSU Marketing Name) The FDR
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Device Serial Number

    Look for the serial numbers MP95A9F6B001, MP95A9F6A001, or MP95A8F5A001 on the device.

How this recall unfolded

Recall timeline

FDA terminated this recall 11 months after the firm started it.

1. Firm started the recall: Jun 15, 201712. FDA classified it Class II: Dec 10, 201723. Published in FDA’s Enforcement Report: Dec 20, 201734. FDA terminated the recall: May 9, 2018420172018
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0232-2018 (June 15, 2017) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0232-2018 to see it in context.
  • Plain-language summary written by qwen3:4b on September 23, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA