GE Inspection Technologies Blade|line Industrial Cabinet X-ray System Recall: Excess

GE Inspection Technologies recalls certain Blade|line Industrial Cabinet X-ray systems due to potential excess radiation emissions under the cabinet, which may exceed safety limits. Affected units have specific serial numbers and were sold between 2013-2014.

🇺🇸 FDA recall #Z-2777-2017Medical deviceGE Inspection Technologies ›
Take action

Contact the company

Recalling firmGE Inspection Technologies, LP50 Industrial Park Rd · Lewistown, PA 17044-9312
✎ GovInfoNow plain-language explainer

What happened, in plain English

GE Inspection Technologies is recalling certain industrial cabinet x-ray systems because they emit radiation levels above the legal safety limit. This could pose a risk to people who access the crawl space under the cabinet, which is normally covered by decorative panels. Read more at govinfonow.com/recall/73545-ge-inspection-technologies-blade-line-industrial-cabinet-x-ray-system.

Source Summary by GovInfoNow · govinfonow.com/recall/73545-ge-inspection-technologies-blade-line-industrial-cabinet-x-ray-system · Updated September 21, 2026 · FDA enforcement record: FDA #Z-2777-2017

Cite this page

“GE Inspection Technologies Blade|line Industrial Cabinet X-ray System Recall: Excess.” GovInfoNow, updated September 21, 2026. https://govinfonow.com/recall/73545-ge-inspection-technologies-blade-line-industrial-cabinet-x-ray-system. Based on FDA recall #Z-2777-2017 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2777-2017%22&limit=1).

Why it’s dangerous

The system emits radiation levels exceeding the legal safety limit of 0.5 milliRoentgen per hour in the crawl space under the cabinet. This space is normally covered by decorative panels and not intended for access, but could be exposed during maintenance or repairs.

Who’s affected

Industrial users of the Blade|line cabinet x-ray systems with specific serial numbers from 2013-2014 are affected. Workers who might access the covered crawl space during maintenance are at higher risk.

How to tell if you have this one

Check for serial numbers listed in the recall: TM 201317, TM 201312, TM 201323, TM 201331, TM 201401, TM 201329, TM 201404, TM 201413, TM 201414, TM 201415, TM 201424, TM 201425, 158017R1. These units were sold between 2013 and 2014.

  • Brand
    GE Inspection Technologies
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Serial Numbers

    Verify if your unit has any of the listed serial numbers: TM 201317, TM 201312, TM 201323, TM 201331, TM 201401, TM 201329, TM 201404, TM 201413, TM 201414, TM 201415, TM 201424, TM 201425, 158017R1.

Where it went and when

Where it was distributed

Distributed in 5 states, and outside the U.S.

Alaska: not listedAKMaine: not listedMEVermont: not listedVTNew Hampshire: not listedNHWashington: not listedWAIdaho: not listedIDMontana: not listedMTNorth Dakota: not listedNDMinnesota: not listedMNIllinois: not listedILWisconsin: not listedWIMichigan: yesMINew York: not listedNYRhode Island: not listedRIMassachusetts: not listedMAOregon: not listedORNevada: not listedNVWyoming: not listedWYSouth Dakota: not listedSDIowa: not listedIAIndiana: not listedINOhio: yesOHPennsylvania: yesPANew Jersey: not listedNJConnecticut: not listedCTCalifornia: not listedCAUtah: not listedUTColorado: not listedCONebraska: not listedNEMissouri: not listedMOKentucky: not listedKYWest Virginia: not listedWVVirginia: not listedVAMaryland: not listedMDDelaware: not listedDEArizona: not listedAZNew Mexico: not listedNMKansas: not listedKSArkansas: not listedARTennessee: not listedTNNorth Carolina: not listedNCSouth Carolina: yesSCDistrict of Columbia: not listedDCOklahoma: not listedOKLouisiana: not listedLAMississippi: not listedMSAlabama: yesALGeorgia: not listedGAHawaii: not listedHITexas: not listedTXFlorida: not listedFL

FDA distribution: “US Distribution to the states of : AL, SC, OH, MI, PA, and Internationally to Germany and Singapore” States read from FDA’s own wording; “not listed” means the notice does not name it.

States: Alabama, Michigan, Ohio, Pennsylvania, South Carolina.

Recall timeline

FDA terminated this recall 2.6 years after the firm started it.

1. Firm started the recall: Jun 28, 201712. FDA classified it Class II: Aug 1, 201723. Published in FDA’s Enforcement Report: Aug 9, 201734. FDA terminated the recall: Feb 6, 2020420172020
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-2777-2017 (June 28, 2017) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-2777-2017 to see it in context.
  • Plain-language summary written by qwen3:4b on September 21, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA