Lower risk Medical device recall · FDA · Published June 29, 2017

Siemens Healthcare ALT Reagent Assays May Give False Drug Results

Siemens Healthcare recalls ALT reagent assays that may cause false high or low drug results for sulfasalazine and sulfapyridine. Affected products include specific part numbers sold by Randox Laboratories.

🇺🇸 FDA recall #Z-1647-2018Medical deviceSiemens Healthcare ›
Take action

Contact the company

Recalling firmSiemens Healthcare Diagnostics, Inc.511 Benedict Ave · Tarrytown, NY 10591-5005
✎ GovInfoNow plain-language explainer

What happened, in plain English

Siemens Healthcare recalls ALT reagent assays that can give false results for sulfasalazine and sulfapyridine drugs. This could lead to incorrect medical decisions if used in blood tests. Read more at govinfonow.com/recall/73498-siemens-healthcare-alt-reagent-assays-may-give-false-drug-results.

Source Summary by GovInfoNow · govinfonow.com/recall/73498-siemens-healthcare-alt-reagent-assays-may-give-false-drug-results · Updated September 20, 2026 · FDA enforcement record: FDA #Z-1647-2018

Cite this page

“Siemens Healthcare ALT Reagent Assays May Give False Drug Results.” GovInfoNow, updated September 20, 2026. https://govinfonow.com/recall/73498-siemens-healthcare-alt-reagent-assays-may-give-false-drug-results. Based on FDA recall #Z-1647-2018 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1647-2018%22&limit=1).

Why it’s dangerous

The reagent assays can interfere with blood tests for sulfasalazine and sulfapyridine, causing false high or low results. This might lead to incorrect medical decisions if the test results are used.

Who’s affected

People using blood tests for sulfasalazine or sulfapyridine treatments may be affected. Those with these medications are most at risk.

How to tell if you have this one

Check for product numbers 10318168, 10309500, 10283341, or 10283342. These reagents were sold by Randox Laboratories and are used for ALT blood tests.

  • Brand
    Siemens Healthcare
  • Category
    Medical device
FDA recall event · 17 products

All 17 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class II
  • Status
    Terminated
  • Distribution
    Distributed in all 50 U.S. states, the District of Columbia, Guam, and Puerto Rico. Distributed in 108 foreign countries: Afghanistan Bolivia Albania¿¿¿¿¿¿¿¿ Algeria¿¿¿¿¿¿¿¿ Angola ARGENTINA Armenia AUSTRALIA Austria¿¿¿¿¿¿¿¿ Azerbaijan Bahamas Bahrain¿¿¿¿¿¿¿¿ Bangladesh¿¿¿¿¿ Belarus¿¿¿¿¿¿¿¿ Belgium¿¿¿¿¿¿¿¿ Bosnia¿Herzeg.¿ Brazil Bulgaria¿¿¿¿¿¿¿ Cambodia Canada Canary islands Chile China Colombia Costa Rica Croatia¿¿¿¿¿¿¿¿ Cyprus¿¿¿¿¿¿¿¿¿ Czech¿Republic¿ Denmark¿¿¿¿¿¿¿¿ Ecuador Egypt¿¿¿¿¿¿¿¿¿¿ El Salavador Estonia¿¿¿¿¿¿¿¿ Finland¿¿¿¿¿¿¿¿ France¿¿¿¿¿¿¿¿¿ Georgia Germany¿¿¿¿¿¿¿¿ Greece¿¿¿¿¿¿¿¿¿ Guatemala Honduras Hong Kong Hungary¿¿¿¿¿¿¿¿ Iceland India Indonesia Iraq Ireland¿¿¿¿¿¿¿¿ Israel¿¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ japan Jordan Kazakhstan¿¿¿¿¿ Korea Kuwait Latvia¿¿¿¿¿¿¿¿¿ Lebanon Lithuania¿¿¿¿¿¿ Luxembourg¿¿¿¿¿ Macedonia Malaysia Mali Malta¿¿¿¿¿¿¿¿¿¿ Mexico Morocco¿¿¿¿¿¿¿¿ Myanmar Nepal Netherlands¿¿¿¿ New Zealand Nicaragua Norway¿¿¿¿¿¿¿¿¿ Oman Pakistan¿¿¿¿¿¿¿ Paraguay Peru Philippines Poland¿¿¿¿¿¿¿¿¿ Portugal¿¿¿¿¿¿¿ Qatar¿¿¿¿¿¿¿¿¿¿ Romania¿¿¿¿¿¿¿¿ Russian¿Fed.¿¿¿ Saudi Arabia Senegal Serbia¿¿¿¿¿¿¿¿¿ Serbia¿¿¿¿¿¿¿¿¿ Singapore Slovakia¿¿¿¿¿¿¿ Slovenia¿¿¿¿¿¿¿ South Africa South Africa South Korea Spain¿¿¿¿¿¿¿¿¿¿ Sri Lanka Sweden¿¿¿¿¿¿¿¿¿ Switzerland¿¿¿¿ Taiwan¿¿¿¿¿¿¿¿¿ Tajikistan Thailand Tunisia¿¿¿¿¿¿¿¿ Turkey¿¿¿¿¿¿¿¿¿ Turkmenistan¿¿¿ United Arab Emerites¿¿¿¿¿¿ United¿Kingdom¿ Uruguay¿¿¿¿¿¿¿¿ Uzbekistan Vatikancity¿¿¿¿ Venezuela Vietnam Yemen
Recall numberProduct, codes & lotsQuantity
Z-1646-2018ADVIA Ammonia (AMM) chemistry assay, an in vitro diagnostic device intended to quantitatively measure Ammonia levels in human plasma (heparin or EDTA). Product Number: 10286035 (Mfr: Randox Laboratories); All lots7014
Z-1647-2018ADVIA Alanine Aminotransferase (ALT) reagent chemistry assays; ALT intended to measure the activity of the enzyme alanine amino transferase (ALT) (also known as a serum glutamic pyruvic transaminase or SGPT) in serum and plasma. (Product Numbers: 10318168, 10309500), ALT P5P (Part Numbers: 10318168/10326245, 10309500/10315181), and ALT Concentrated (Part Numbers: 10283341, 10283341, 10283342); (Mfr: Randox Laboratories); All lots33,233
Z-1648-2018ADVIA Salicylate (SAL) reagent chemistry assay; use in the quantitative determination of salicylate in human serum and plasma (lithium heparin). Product Number: 10327382 (Mfr: Sekisui Diagnostics); All lots1667
Z-1649-2018Dimension Ammonia (AMM) assay; (Product Number: DF119/10711991) (Mfr: Siemens Healthcare Diagnostics, Newark, DE); All lots37,923
Z-1650-2018Dimension Alanine Aminotransferase (ALTI) assay; (Product Number: DF143/10475530) (Mfr: Siemens Healthcare Diagnostics, Newark, DE); All lots276,511
Z-1651-2018Dimension Aspartate Aminotransferase (AST) assay (Product Number: DF41A/10444959) (Mfr: Siemens Healthcare Diagnostics, Newark, DE); All lots181,072
Z-1652-2018Dimension Glucose (GLUC) assay (Product Number: DF40/10444973) (Mfr: Siemens Healthcare Diagnostics, Newark, DE); All lots80,466
Z-1653-2018Dimension Creatine Kinase MB (MBI) assay (Product Number: DF32/10464510) (Mfr: Siemens Healthcare Diagnostics, Newark, DE); All lots30,446
Z-1654-2018Dimension Thyroxine (T4) assay (Product Number: DF65/10444908) (Mfr: Siemens Healthcare Diagnostics, Newark, DE); All lots10,056
Z-1655-2018Dimension Vista Ammonia (AMM) assay (Product Number: K3119/10711992) (Mfr: Siemens Healthcare Diagnostics, Newark, DE); All lots34,590
Z-1656-2018Dimension Vista Alanine Aminotransferase (ALTI) assay; (Product Number: K2143/10635565) (Mfr: Siemens Healthcare Diagnostics, Newark, DE); All lots109,334
Z-1657-2018Dimension Vista Aspartate Aminotransferase (AST) assay; (Product Number: K2014/10445148) (Mfr: Siemens Healthcare Diagnostics, Newark, DE); All lots32,744
Z-1658-2018Dimension Vista Glucose (GLU) assay; (Product Number: K1039/10445162) (Mfr: Siemens Healthcare Diagnostics, Newark, DE); All lots19,245
Z-1659-2018Dimension Vista Creatine Kinase MB (MBI) assay; (Product Number: K3032/10464339) (Mfr: Siemens Healthcare Diagnostics, Newark, DE); All lots3130
Z-1660-2018Dimension Vista Thyroxine (T4) assay; (Product Number: K6065/10445101) (Mfr: Siemens Healthcare Diagnostics, Newark, DE); All lots19,991
Z-1661-2018Syva Emit 2000 Tacrolimus Assay (Product Number: 8R019UL/10445397) (Mfr: Siemens Healthcare Diagnostics, Newark, DE); All lots10,067
Z-1662-2018Syva Emit 2000 Sirolimus Assay (Product Number: 8S019UL/10445401) (Mfr: Siemens Healthcare Diagnostics, Newark, DE); All lots226
Based on the official FDA remedy

What to do — step by step

  1. Check Product Numbers

    Look for product numbers 10318168, 10309500, 10283341, or 10283342 on the reagent packaging.

Where it went and when

Where it was distributed

Distributed nationwide, and outside the U.S.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFLPuerto Rico: yesPR

FDA distribution: “Distributed in all 50 U.S. states, the District of Columbia, Guam, and Puerto Rico. Distributed in 108 foreign countries: Afghanistan Bolivia Albania¿¿¿¿¿¿¿¿ Algeria¿¿¿¿¿¿¿¿ Angola ARGENTINA Armenia AUSTRALIA Austria¿¿¿¿¿¿¿¿ Azerbaijan…” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide, and outside the U.S.

Recall timeline

FDA terminated this recall 2.8 years after the firm started it.

1. Firm started the recall: Jun 29, 201712. FDA classified it Class II: May 4, 201823. Published in FDA’s Enforcement Report: May 16, 201834. FDA terminated the recall: Apr 23, 2020420172020
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

✎ GovInfoNow

Questions people ask

Do I need to stop using these reagents?

No, but you should check if you have the affected product numbers. The recall does not require immediate stop use.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1647-2018 (June 29, 2017) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1647-2018 to see it in context.
  • Plain-language summary written by qwen3:4b on September 20, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA