Skip to content
Lower risk Medical device recall · FDA · Published July 28, 2017

Roche Cobas 8000 System Software Version 05-02 and 06-03 May Cause Incorrect Test Results

Roche recalls Cobas 8000 Modular Series systems with specific software versions due to potential incorrect test results. Affected units have unique serial numbers listed in the recall notice.

🇺🇸 FDA recall #Z-0675-2018Medical deviceRoche ›
Take action

Contact the company

Recalling firmRoche Diagnostics Corporation9115 Hague Rd · Indianapolis, IN 46256-1025
✎ GovInfoNow plain-language explainer

What happened, in plain English

Roche is recalling certain Cobas 8000 Modular Series systems with software versions 05-02 and 06-03 because a software failure might set the system to default settings, leading to incorrect test results. This could affect the accuracy of medical tests. Read more at govinfonow.com/recall/73332-roche-cobas-8000-system-software-version-05-02-and-06.

Source Summary by GovInfoNow · govinfonow.com/recall/73332-roche-cobas-8000-system-software-version-05-02-and-06 · Updated September 20, 2026 · FDA enforcement record: FDA #Z-0675-2018

Cite this page

“Roche Cobas 8000 System Software Version 05-02 and 06-03 May Cause Incorrect Test Results.” GovInfoNow, updated September 20, 2026. https://govinfonow.com/recall/73332-roche-cobas-8000-system-software-version-05-02-and-06. Based on FDA recall #Z-0675-2018 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0675-2018%22&limit=1).

Why it’s dangerous

The software failure may reset the system to default settings, causing the device to produce incorrect test results. This could lead to misdiagnosis or inappropriate treatment decisions if the results are used for patient care.

Who’s affected

Healthcare facilities using the Cobas 8000 Modular Series system with the specified software versions are most at risk. Patients relying on these systems for medical testing could receive inaccurate results.

How to tell if you have this one

Check if your Cobas 8000 Modular Series system has software versions 05-02 or 06-03. The recall includes specific serial numbers listed in the notice.

  • Brand
    Roche
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Software Version

    Verify your system's software version by checking the device settings or documentation. If it is 05-02 or 06-03, it is affected.

Where it went and when

Where it was distributed

Distributed in 36 states and Puerto Rico.

Alaska: not listedAKMaine: not listedMEVermont: not listedVTNew Hampshire: yesNHWashington: yesWAIdaho: not listedIDMontana: not listedMTNorth Dakota: not listedNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: not listedNVWyoming: not listedWYSouth Dakota: not listedSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: not listedWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: not listedNMKansas: yesKSArkansas: not listedARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: not listedSCDistrict of Columbia: not listedDCOklahoma: yesOKLouisiana: yesLAMississippi: not listedMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFLPuerto Rico: yesPR

FDA distribution: “Distributed in 36 states: AL, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, NC, NE, NH, NJ, NY, OH, OK, OR, PA, RI, TN, TX, UT, VA, WA, WI and Puerto Rico.” States read from FDA’s own wording; “not listed” means the notice does not name it.

States: Alabama, Arizona, California, Colorado, Connecticut, Delaware, Florida, Georgia, Hawaii, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maryland, Massachusetts, Michigan, Minnesota, Missouri, Nebraska, New Hampshire, New Jersey, New York, North Carolina, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, Rhode Island, Tennessee, Texas, Utah, Virginia, Washington, Wisconsin.

Recall timeline

FDA terminated this recall 2.9 years after the firm started it.

1. Firm started the recall: Jul 28, 201712. FDA classified it Class II: Feb 21, 201823. Published in FDA’s Enforcement Report: Feb 28, 201834. FDA terminated the recall: Jun 10, 2020420172020
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

✎ GovInfoNow

Questions people ask

Do I need to stop using my Cobas 8000 system immediately?

No, the recall does not require immediate discontinuation. You can continue using the system but should check the software version and follow the action steps.

What should I do if my system has the affected software versions?

Check your system's serial number against the list provided in the recall notice. If it matches, contact Roche for further instructions.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0675-2018 (July 28, 2017) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0675-2018 to see it in context.
  • Plain-language summary written by qwen3:4b on September 20, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA