Skip to content
Use caution Medical device recall · FDA · Published September 19, 2017

Intuitive Surgical daVinci X EndoWrist Stapler 45 Recall: Incorrect Wrench Interface May

Intuitive Surgical recalls daVinci X EndoWrist Stapler 45 models with serial numbers S10170630, S11170705, S10170707, S10170731 due to a potential manufacturing defect that may prevent the Stapler Release Kit Wrench from unclamping the instrument.

🇺🇸 FDA recall #Z-0404-2018Medical deviceIntuitive Surgical ›
Take action

Contact the company

Recalling firmIntuitive Surgical, Inc.1266 Kifer Rd Bldg 101 · Sunnyvale, CA 94086-5304
✎ GovInfoNow plain-language explainer

What happened, in plain English

Intuitive Surgical is recalling specific daVinci X EndoWrist Stapler 45 instruments that may have a manufacturing defect causing the Stapler Release Kit Wrench to not unclamp the instrument. This could lead to complications during surgery if the instrument cannot be properly released. Read more at govinfonow.com/recall/73006-intuitive-surgical-davinci-x-endowrist-stapler-45-recall-incorrect-wrench.

Source Summary by GovInfoNow · govinfonow.com/recall/73006-intuitive-surgical-davinci-x-endowrist-stapler-45-recall-incorrect-wrench · Updated September 17, 2026 · FDA enforcement record: FDA #Z-0404-2018

Cite this page

“Intuitive Surgical daVinci X EndoWrist Stapler 45 Recall: Incorrect Wrench Interface May.” GovInfoNow, updated September 17, 2026. https://govinfonow.com/recall/73006-intuitive-surgical-davinci-x-endowrist-stapler-45-recall-incorrect-wrench. Based on FDA recall #Z-0404-2018 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0404-2018%22&limit=1).

Why it’s dangerous

The Stapler Release Kit Wrench interface feature accessed through release hole 2 may be manufactured to an incorrect dimension. This could prevent the wrench from manually unclamping the instrument during surgery, potentially causing complications.

Who’s affected

Surgical teams using the daVinci X EndoWrist Stapler 45 with serial numbers S10170630, S11170705, S10170707, or S10170731 are most at risk. These instruments are used in general, thoracic, gynecologic, and urologic surgeries.

How to tell if you have this one

Check for serial numbers S10170630, S11170705, S10170707, or S10170731 on the daVinci X EndoWrist Stapler 45 instruments. These models were sold with the da Vinci Surgical System (Model IS4000) for specific surgical procedures.

  • Brand
    Intuitive Surgical
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Serial Numbers

    Verify if your instrument has serial numbers S10170630, S11170705, S10170707, or S10170731.

Where it went and when

Where it was distributed

Distributed nationwide.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “US Nationwide Distribution” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide.

Recall timeline

FDA terminated this recall 6.4 years after the firm started it.

1. Firm started the recall: Sep 19, 201712. FDA classified it Class II: Jan 23, 201823. Published in FDA’s Enforcement Report: Jan 31, 201834. FDA terminated the recall: Jan 30, 2024420172024
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0404-2018 (September 19, 2017) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0404-2018 to see it in context.
  • Plain-language summary written by qwen3:4b on September 17, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA