Lower risk Medical device recall · FDA · Published October 4, 2017

Draegar Medical Systems: Infinity Dual Hemo MCable Pod Recall - Liquid Ingress Risk

Draegar Medical Systems recalls Infinity Dual Hemo MCable Pods with Revision Indexes 15 or 16 due to potential liquid ingress causing inaccurate blood pressure readings. Affected units may be sold in medical settings.

🇺🇸 FDA recall #Z-0942-2018Medical deviceDraeg: Medical Systems ›
Take action

Contact the company

Recalling firmDraegar Medical Systems, Inc.6 Tech Dr · Andover, MA 01810-2434
✎ GovInfoNow plain-language explainer

What happened, in plain English

Draegar Medical Systems is recalling Infinity Dual Hemo MCable Pods with Revision Indexes 15 or 16. These pods may allow liquid to enter, leading to incorrect blood pressure measurements in medical devices. Read more at govinfonow.com/recall/72912-draegar-medical-systems-infinity-dual-hemo-mcable-pod-recall-liquid.

Source Summary by GovInfoNow · govinfonow.com/recall/72912-draegar-medical-systems-infinity-dual-hemo-mcable-pod-recall-liquid · Updated September 17, 2026 · FDA enforcement record: FDA #Z-0942-2018

Cite this page

“Draegar Medical Systems: Infinity Dual Hemo MCable Pod Recall - Liquid Ingress Risk.” GovInfoNow, updated September 17, 2026. https://govinfonow.com/recall/72912-draegar-medical-systems-infinity-dual-hemo-mcable-pod-recall-liquid. Based on FDA recall #Z-0942-2018 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0942-2018%22&limit=1).

Why it’s dangerous

The pods may let liquid enter despite their IPX4 rating, which is designed to prevent water ingress. This could cause inaccurate blood pressure readings, potentially leading to incorrect medical decisions.

Who’s affected

Medical professionals using the Infinity Acute Care System with the affected revision indexes are most at risk. Patients receiving blood pressure monitoring via this device could be impacted.

How to tell if you have this one

Check the product label for Revision Indexes 15 or 16. The affected pods are used with the Dr¿ger Infinity Acute Care System.

  • Brand
    Draeg: Medical Systems
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Product Label

    Look for Revision Indexes 15 or 16 on the product label to identify affected units.

Where it went and when

Where it was distributed

Distributed nationwide.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “Nationally” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide.

Recall timeline

FDA terminated this recall 8 months after the firm started it.

1. Firm started the recall: Oct 4, 201712. FDA classified it Class II: Mar 9, 201823. Published in FDA’s Enforcement Report: Mar 21, 201834. FDA terminated the recall: Jun 15, 2018420172018
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0942-2018 (October 4, 2017) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0942-2018 to see it in context.
  • Plain-language summary written by qwen3:4b on September 17, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA