Skip to content
Unrated Medical device recall · FDA · Published October 19, 2017

Philips CombiDiagnost R9: X-ray System Recall - Tube Adapter Risk

Philips recalls CombiDiagnost R90 X-ray systems where tube adapter-plate wiring was improperly installed, risking electrical hazards. Affected units may need repair or replacement.

🇺🇸 FDA recall #Z-0330-2018Medical devicePhilips ›
Take action

Contact the company

Recalling firmPhilips Electronics North America Corporation3000 Minuteman Rd · Andover, MA 01810-1032
✎ GovInfoNow plain-language explainer

What happened, in plain English

Philips is recalling CombiDiagnost R90 X-ray systems that may have had the tube adapter-plate green/yellow protective earthing wire removed and spacers used incorrectly. This could cause electrical hazards if the system is used without proper grounding. Read more at govinfonow.com/recall/72809-philips-combidiagnost-r9-x-ray-system-recall-tube-adapter-risk.

Source Summary by GovInfoNow · govinfonow.com/recall/72809-philips-combidiagnost-r9-x-ray-system-recall-tube-adapter-risk · Updated September 14, 2026 · FDA enforcement record: FDA #Z-0330-2018

Cite this page

“Philips CombiDiagnost R9: X-ray System Recall - Tube Adapter Risk.” GovInfoNow, updated September 14, 2026. https://govinfonow.com/recall/72809-philips-combidiagnost-r9-x-ray-system-recall-tube-adapter-risk. Based on FDA recall #Z-0330-2018 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0330-2018%22&limit=1).

Why it’s dangerous

The tube adapter-plate green/yellow protective earthing wire was removed, and spacers with different thickness were used to position the pre-loc ring. This improper installation could lead to electrical hazards if the system is used without proper grounding.

Who’s affected

Healthcare facilities using Philips CombiDiagnost R90 X-ray systems are likely affected. Medical staff and patients using these systems could be at risk if the electrical issue remains uncorrected.

How to tell if you have this one

Check for CombiDiagnost R90 X-ray systems with UDI 709030. The issue involves improper installation of the tube adapter-plate wiring and spacers.

  • Brand
    Philips
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check for Improper Wiring

    Inspect the tube adapter-plate for a removed green/yellow protective earthing wire and incorrect spacer use.

How this recall unfolded

Recall timeline

FDA terminated this recall 11 months after the firm started it.

1. Firm started the recall: Oct 19, 201712. FDA classified it Class II: Jan 10, 201823. Published in FDA’s Enforcement Report: Jan 17, 201834. FDA terminated the recall: Sep 20, 2018420172018
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0330-2018 (October 19, 2017) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0330-2018 to see it in context.
  • Plain-language summary written by qwen3:4b on September 14, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA