Philips recalls Brilliance iCT CT scanners made from March 2009 onward due to potential loose screws in the gantry that could cause noise but not immediate danger.
There is a possibility that screws holding an internal portion of the gantry (Reaction Ring) to the scanner s main bearing may come loose. If this occurs, the part(s) will not be expelled, however, a noise may be heard coming from the gantry.
Philips is recalling its Brilliance iCT CT scanners made from March 5, 2009, onward. Loose screws in the gantry could cause noise but do not pose an immediate risk of injury or malfunction.
Screws holding the gantry to the scanner's main bearing could become loose over time. If loose, they might cause a noise but the part won't expel itself, meaning no immediate injury risk.
Healthcare facilities using Philips Brilliance iCT CT scanners made from March 5, 2009, onward are likely affected. Staff and patients using these scanners may be at risk if the gantry makes unusual noises.
Check for Philips Healthcare Brilliance iCT CT scanners with model number 728306. These were manufactured from March 5, 2009, through present.
Listen for noise from the gantry during operation; if heard, stop using the scanner immediately.
The recall states there is a possibility of loose screws causing noise but no immediate injury risk. The hazard is low risk.
The scanners were manufactured from March 5, 2009, through present.
Stop using the scanner immediately and contact Philips for further instructions.
We’ve drafted a message you can send Philips to request your refund or repair — edit it as you like.
Subject: Recall Z-0331-2018 — Philips Philips Hello, I own a Philips that is covered by recall Z-0331-2018 from Philips. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.