Skip to content
Use caution Medical device recall · FDA · Published July 8, 2026

Tobii Dynavox TD Navio Maxi Speech Device Recall: Battery Short Circuit Risk

Tobii Dynavox recalls TD Navio Maxi speech devices with model 1000799 due to potential battery short circuits causing heat and gas. Affected units have specific serial numbers and were sold from 2024 to 2026.

🇺🇸 FDA recall #Z-2960-2026Medical deviceTobii Dynavox ›
✎ GovInfoNow plain-language explainer

Tobii Dynavox is recalling TD Navio Maxi speech devices with model number 1000799 because damaged battery mounting tabs could cause internal short circuits, leading to heat and gas venting. This risk is serious enough to require immediate action to prevent potential device damage or injury. Read more at govinfonow.com/recall/697932-tobii-dynavox-td-navio-maxi-speech-device-recall-battery-short.

Source Summary by GovInfoNow · govinfonow.com/recall/697932-tobii-dynavox-td-navio-maxi-speech-device-recall-battery-short · Updated September 14, 2026 · FDA enforcement record: FDA #Z-2960-2026

Cite this page

“Tobii Dynavox TD Navio Maxi Speech Device Recall: Battery Short Circuit Risk.” GovInfoNow, updated September 14, 2026. https://govinfonow.com/recall/697932-tobii-dynavox-td-navio-maxi-speech-device-recall-battery-short. Based on FDA recall #Z-2960-2026 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2960-2026%22&limit=1).

The FDA announced on July 8, 2026, that Tobii Dynavox Llc is voluntarily recalling the TD Navio Maxi Speech Generating Device due to a potential battery mounting tab failure. This defect may cause localized heating and gas venting within the device enclosure.

How to tell if you have it

  • Brand: Tobii Dynavox
  • Product: TD Navio Maxi Speech Generating Device
  • Model/Catalog Number: 1000799
  • UDI-DI: 073400747602070
  • Serial Numbers: TDNL1-240904000016 through TDNL1-260617070001

The hazard

Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure. This movement potentially leads to an internal short circuit, which may cause localized heating and gas venting within the device enclosure.

Context

GovInfoNow's records show Tobii Dynavox Llc has one other recall on record since February 24, 2023. On February 24, 2023, the company recalled the TD I-110 Speech Device because the battery may come loose from the housing and become damaged. Other recent recalls of similar products include the Nerivio Infinity Software Version 4.1.2.0 Recall (March 23, 2026) and the Philips Allura Xper FD10C Medical System Recall (December 15, 2025).

How to tell if you have this one

Check for the model number 1000799 on the device. Affected units have serial numbers starting with TDNL1-240904000016 and ending with TDNL1-260617070001. The device was sold between 2024 and 2026.

  • Brand
    Tobii Dynavox
  • Category
    Medical device
FDA recall event · 2 products

All 2 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class II
  • Status
    Ongoing
  • Distribution
    US Nationwide distribution.
Recall numberProduct, codes & lotsQuantity
Z-2959-2026TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798; UDI-DI 073400747602056 Serial Numbers TDNM1-240628000004 through TDNM1-260617070001. Corrected/future production is differentiated by new serialization from TDNM1-260617070001 onward (production batch 20260615, improved design).—
Z-2960-2026TD Navio Maxi Speech Generating Device Model/Catalog Number: 1000799; UDI-DI 073400747602070 Serial Numbers TDNL1-240904000016 through TDNL1-260617070001. Corrected/future production is differentiated by new serialization from TDNL1-260617070001 onward (production batch 20260615, improved design).—
Based on the official FDA remedy

What to do — step by step

  1. Check Device Serial Number

    Look for the serial number on the device, which should start with TDNL1-240904000016 and end with TDNL1-260617070001.

  2. Contact Manufacturer

    Reach out to Tobii Dynavox for further instructions on how to address the issue.

Where it went and when

Where it was distributed

Distributed nationwide.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “US Nationwide distribution.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide.

Recall timeline

FDA published this recall 49 days after the firm started it.

1. Firm started the recall: Jul 8, 202612. FDA classified it Class II: Aug 14, 202623. Published in FDA’s Enforcement Report: Aug 26, 202634. Added to GovInfoNow: Sep 8, 202642026
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. Added to GovInfoNow

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-2960-2026 (July 8, 2026) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-2960-2026 to see it in context.
  • Plain-language summary written by qwen3:4b on September 14, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA