Pacific Medical Group Inc. recalled Ultrasound Transducers (M1356A) serviced with non-OEM components, the FDA said on Sept. 11, 2019. The agency announced the voluntary Class II recall due to risks of inaccurate fetal heart rate detection and patient burns.
How to tell if you have it
- Brand: Pacific Medical
- Product: Ultrasound Transducer (M1356A)
- Serviced with: Pacific Medical Philips M1356A Ultrasound Board (NFPH9510), Bottom Case (NFPH9515), Cable Assembly (NFPH9520), or Crystal (NFPH9535)
- Serial numbers serviced between 12/18/14 and 4/2/19
- Units affected: 1994
The hazard
The devices were serviced or remanufactured using non-OEM equivalent components that have not been appropriately verified or validated. This may cause inability to accurately detect and measure fetal heart rate, maternal peripheral oxygen saturation, uterine activity, temperature, and blood pressure, lack of electrocardiography tracing, patient/clinician burns, and delay in detecting maternal or fetal distress.
Context
GovInfoNow's records show Pacific Medical Group Inc. has two other recalls on record since Sept. 19, 2019. On March 24, 2021, the company recalled Aftermarket Alaris Infusion Pump Module Model 8100 Front Bezel due to risk of free flow or under infusion. On Sept. 19, 2019, Pacific Medical Group recalled GE Corometrics Ultrasound Transducers for distribution before 510(k) clearance. Other recent recalls of similar products include Philips Avalon Fetal Monitor Recall: Incorrect Speaker Connector Risk (May 26, 2026) and LH One Step Ovulation Test Device Recall: False Results Risk Medical Errors (Nov. 22, 2025).