Every recall we track for Pacific Medical Group Inc. — 18 in all — plus the complaints and injury reports the public filed about them.
Pacific Medical Group recalls Aftermarket Alaris Infusion Pump Module Model 8100 Front Bezel components installed by their service team. Over time, posts may crack or separate, leading to free flow, over infusion, or under infusion. Affected units have serial numbers from 07/10/2020 to 02/18/2021.
Pacific Medical recalls GE Corometrics Nautilus Ultrasound 5700 (Old Style) models sold between Jan 2015 and Apr 2019 due to fetal transducers distributed without required 510(k) clearance. Affected units have serial numbers from 01/21/2015 to 04/30/2019.
Pacific Medical recalls certain fetal transducers sold between Feb 2016 and Apr 2019 that were distributed without required U.S. regulatory clearance. This affects pregnant women using the device for monitoring.
Pacific Medical recalls GE Corometrics Toco Transducers (2264LAX) made between Feb 2016 and Apr 2019 due to potential fetal monitoring inaccuracies. Affected units may not detect fetal heart rate or maternal oxygen levels correctly, risking undetected distress.
Pacific Medical recalls Philips M1356A ultrasound transducers sold between Jan 2015 and Apr 2019 due to distribution before required 510(k) clearance. This may pose regulatory risks but not immediate safety hazards.
Pacific Medical recalls Toco Transducers (M1355A) sold between Oct 2014 and Apr 2019 that were distributed without required U.S. regulatory clearance for fetal monitoring.
Pacific Medical recalls GE Corometrics Nautilus Ultrasound Transducers (5700HAX) sold between Jan 2015 and Apr 2019 due to distribution before required 510(k) clearance for fetal use. Affected units may pose risks if used without proper regulatory approval.
Pacific Medical recalls GE Corometrics Nautilus Ultrasound Transducer models 5700LAX (FFCM5260) sold between Jan 2015 and Apr 2019 due to distribution before required 510(k) clearance for fetal use.
Pacific Medical recalls ultrasound transducer M2736A that may have been serviced with non-OEM parts. This could cause inaccurate fetal heart rate, oxygen levels, and uterine activity monitoring. Affected units were serviced between September 2015 and April 2019.
Pacific Medical recalls Button-Style Tocotransducers (Nautilus) with serial numbers serviced between April 2015 and April 2019 that may not accurately monitor fetal heart rate, oxygen levels, or contractions due to improper servicing.
Pacific Medical recalls Toco+ transducers with ECG/IUP capability that may have been serviced with non-OEM components. These devices could fail to accurately monitor fetal heart rate, oxygen levels, contractions, and other critical maternal health indicators.
Pacific Medical recalls Toco Transducer M1355A models serviced with non-OEM parts between Oct 2014 and Apr 2019. These devices may fail to monitor fetal heart rate, oxygen levels, and contractions, risking maternal and fetal health.
Pacific Medical recalls Nautilus ultrasound transducers (model 5700LAX) that may have been serviced with non-OEM parts. These devices could fail to monitor fetal heart rate, oxygen levels, contractions, or cause patient burns. Affected units were serviced between Dec 2014 and Apr 2019.
Pacific Medical recalls Toco M2734A fetal monitors that may have been serviced with non-OEM cables. These devices could fail to accurately track fetal heart rate, oxygen levels, and contractions, risking maternal and fetal health.
Pacific Medical recalls Nautilus ultrasound transducers (model 5700HAX) that may have been serviced with non-OEM parts between Oct 2014 and Apr 2019. These devices could fail to accurately monitor fetal heart rate, oxygen levels, contractions, and other critical health indicators during pregnancy.
Pacific Medical recalls Toco MP Transducers (M2734B) serviced with non-OEM parts between Sept 2015 and Apr 2019. These devices may fail to detect fetal heart rate, oxygen levels, or contractions, risking maternal/fetal distress.
Pacific Medical recalls Loop-Style Tocotransducer 2264LAX units serviced with non-OEM parts between Jan 2015 and Apr 2019. These devices may fail to monitor fetal heart rate, oxygen levels, contractions, and other critical health indicators during pregnancy.
Pacific Medical recalls ultrasound transducers M1356A that may have been serviced with non-OEM parts between Dec 2014 and Apr 2019. These devices could fail to accurately monitor fetal heart rate, oxygen levels, contractions, and maternal health, risking undetected distress.