LivaNova recalls CP5 Drive Units (serials 60E03336-60E03347) for potential missing ferrite ring causing electrical issues in heart surgery equipment.
One ferrite ring expected to be installed on centrifugal pump system power cable may not be present.
LivaNova is recalling specific CP5 Drive Units used in heart surgery systems due to a possible missing ferrite ring that could cause electrical problems. This affects only certain units with serial numbers starting with 60E03336 to 60E03347.
The CP5 Drive Unit may have a ferrite ring missing from the power cable, which could lead to electrical interference during heart surgery. This might cause equipment malfunctions but does not pose an immediate danger to patients.
Heart surgery teams using the specific CP5 Drive Units with serial numbers 60E03336 to 60E03347 are most at risk. These units were sold in the U.S. for use with HLM S5/C5 systems.
Check if your CP5 Drive Unit has a serial number starting with 60E03336, 60E03339, 60E03344, or 60E03347. These units were used in HLM S5/C5 heart surgery systems.
Verify your CP5 Drive Unit's serial number to confirm if it matches the affected range: 60E03336, 60E03339, 60E03344, or 60E03347.
The recall is due to a possible missing ferrite ring on the power cable, which could cause electrical interference during heart surgery.
No, the recall does not require immediate stoppage. Affected units should be checked for the missing ferrite ring.
Check your CP5 Drive Unit's serial number to see if it starts with 60E03336, 60E03339, 60E03344, or 60E03347.
We’ve drafted a message you can send LivaNova to request your refund or repair — edit it as you like.
Subject: Recall Z-0787-2022 — LivaNova LivaNova Hello, I own a LivaNova that is covered by recall Z-0787-2022 from LivaNova. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.