Every recall we track for LivaNova Deutschland GmbH — 5 in all — plus the complaints and injury reports the public filed about them.
LivaNova recalls Essenz HLM heart-lung machines with software version 1.5 due to touchscreen blackouts that may cause hypoxia. Affected units have specific serial numbers and were sold in Germany.
LivaNova recalls 3T Heater Cooler Systems with incorrect maintenance labeling in Quick Reference Guides. Affected units may require correction to prevent improper maintenance.
LivaNova recalls CP5 Drive Units (serials 60E03336-60E03347) for potential missing ferrite ring causing electrical issues in heart surgery equipment.
LivaNova recalls SCP Pump Control Panel S5/C5 models with serial numbers 60S14735-60S14755. Premature angle encoder wear may cause stroke, organ damage, or tissue injury during heart surgery. Stop using immediately.
LivaNova recalls SCP Pump Control Panel REF 60-02-15 due to potential angle encoder failure causing stroke, organ damage, or arterial injury during heart surgery. Affected units have specific serial numbers and UDI codes.