The FDA announced on Oct. 18, 2024, that LivaNova Deutschland GmbH is recalling Essenz HLM Heart-Lung Machines due to a touchscreen hazard. The agency issued recall number Z-0571-2025 for the voluntary Class II action affecting 38 units distributed nationwide.
How to tell if you have it
- Brand: LivaNova
- Product: Essenz HLM (Heart-Lung Machine)
- REF: 49-00-10
- Software Version: 1.5
- Serial Numbers: 49PK00436, 49PK00108, 49PK00374, 49PK00479, 49PK00480, 49PK00481, 49PK00093, 49PK00091, 49PK00092, 49PK00163, 49PK00497, 49PK00498, 49PK00219, 49PK00223, 49PK00224, 49PK00225, 49PK00226, 49PK00241, 49PK00255, 49PK00292, 49PK00291, 49PK00293, 49PK00490, 49PK00487, 49PK00488, 49PK00278, 49PK00274, 49PK00277, 49PK00275, 49PK00276, 49PK00273, 49PK00457, 49PK00458, 49PK00459, 49PK00455, 49PK00456, 49PK00290, 49PK00288
The hazard
According to the FDA, the Essenz HLM Heart-Lung Machine's GUI touchscreen may go dark for 10-15 seconds while safety functions remain operational. A second reset may cause the gas blender to enter standby mode and deactivate sensors until a new case is started. This reset can lead to hypoxia and hypoperfusion.
Context
GovInfoNow's records show LivaNova Deutschland GmbH has issued three other recalls since Aug. 26, 2019. On Dec. 17, 2021, the firm recalled CP5 Drive Units due to a missing ferrite ring risk. Other recent recalls of similar products include Medline Oxygen Masks Recall (Nov. 21, 2025) and Baxter Blue Ventilator Adapter Module Recall (Feb. 18, 2026).