Use caution Medical device recall · FDA · Published June 27, 2022

Hamilton C6 Ventilator Recall: Loose Indicator Board May Cause Water Ingress

Hamilton Medical recalls C6 ventilators with serial numbers starting 5663-9590 due to loose status indicator board risking water ingress from disinfectants, potentially causing technical faults and unsafe breathing conditions.

🇺🇸 FDA recall #Z-1525-2022Medical deviceHamilton Medical ›
Take action

Contact the company

Recalling firmHamilton Medical AGVia Crusch 8 · Bonaduz, Switzerland
✎ GovInfoNow plain-language explainer

Hamilton Medical is recalling specific C6 intensive care ventilators because the status indicator board can become loose, allowing water from disinfectants to enter. This may cause repeated technical faults that could force the ventilator to switch to unsafe breathing modes, risking patient safety. Read more at govinfonow.com/recall/60419-hamilton-c6-ventilator-recall-loose-indicator-board-may-cause-water.

Source Summary by GovInfoNow · govinfonow.com/recall/60419-hamilton-c6-ventilator-recall-loose-indicator-board-may-cause-water · Updated August 19, 2026 · FDA enforcement record: FDA #Z-1525-2022

Cite this page

“Hamilton C6 Ventilator Recall: Loose Indicator Board May Cause Water Ingress.” GovInfoNow, updated August 19, 2026. https://govinfonow.com/recall/60419-hamilton-c6-ventilator-recall-loose-indicator-board-may-cause-water. Based on FDA recall #Z-1525-2022 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-1525-2022%22&limit=1).

The FDA announced on June 27, 2022, that Hamilton Medical AG is recalling HAMILTON-C6 Intensive Care Ventilators because a loose status indicator board can allow water ingress, potentially causing technical faults or unsafe ventilation modes. Owners should contact the company for repairs.

How to tell if you have it

  • Brand: Hamilton Medical
  • Product: HAMILTON-C6 Intensive Care Ventilator
  • REF: 160021
  • UDI-DI: 07630002808590
  • Units affected: 358

The hazard

The ventilator status indicator board can become loose, which could lead to water ingress from disinfectants that may cause technical fault alarms. Multiple technical faults in a short time may force ventilators into Safety Ventilation or Ambient State with a Panel connection lost message displayed.

Context

GovInfoNow's records show Hamilton Medical AG has 11 other recalls on record since March 9, 2018. Other recent recalls of similar products include the Laborie Solar GI System Recall (May 21, 2026), the Baxter Blue Ventilator Adapter Module Recall (Feb. 18, 2026), and the ICU Medical Hotline Fluid Warming Set Recall (June 24, 2026).

How to tell if you have this one

Check the serial numbers listed in the recall notice (5663 to 9590). The ventilator model is the Hamilton C6 Intensive Care Ventilator with reference number 160021.

  • Brand
    Hamilton Medical
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Ventilator Status

    Inspect the status indicator board for looseness or signs of water damage.

  2. Contact Manufacturer

    Reach out to Hamilton Medical for guidance on next steps if the indicator board appears loose.

Where it went and when

Where it was distributed

Distributed nationwide.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFLPuerto Rico: yesPR

FDA distribution: “US nationwide distribution including Puerto Rico.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide.

Recall timeline

FDA published this recall 2 months after the firm started it.

1. Firm started the recall: Jun 27, 202212. FDA classified it Class I: Aug 19, 202223. Published in FDA’s Enforcement Report: Aug 31, 202232022
  1. Firm started the recall
  2. FDA classified it Class I
  3. Published in FDA’s Enforcement Report

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-1525-2022 (June 27, 2022) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-1525-2022 to see it in context.
  • Plain-language summary written by qwen3:4b on August 19, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA