The FDA announced on May 15, 2024, that Hamilton Medical AG is voluntarily recalling the HAMILTON-C6 ventilator (PN: 160021) due to a Class I hazard. The device may fail to re-initiate ventilation after a sensor error, potentially causing patient hypoxia.
How to tell if you have it
- Brand: Hamilton Medical
- Product: Ventilator HAMILTON-C6
- Part Number: PN: 160021
- UDI: 07630002808590
- Software Versions: SW v1.1.4, SW v1.1.5, SW v1.1.6
The hazard
The ventilator may enter a sensor fail mode where ventilation does not re-initiate after the patient is reconnected following an O2 enrichment and tube disconnection sequence. This failure can cause hypoxia, according to the FDA record.
Context
GovInfoNow's records show Hamilton Medical AG has 11 other recalls on record since March 9, 2018. Recent brand recalls include a breathing circuit set risk dated May 29, 2026, and an IntelliCuff leakage error dated May 22, 2026. Other recent recalls of similar products include the Baxter Blue Ventilator Adapter Module Recall (Feb. 18, 2026) and the Draeger Atlan A350 Anesthesia Workstation Recall (May 7, 2026).