Philips recalls MR 7700 systems with specific software versions due to potential stiffness value errors in exported elastography maps. Affected units may show inaccurate medical imaging data.
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).
Philips is recalling certain MR 7700 systems that may display incorrect stiffness values in elastography images. This could lead to inaccurate medical diagnoses if the images are used for treatment decisions.
The software versions R11.1 and R12.1 may cause stiffness value errors when exporting elastography images to PACS systems. This could result in inaccurate medical imaging data being used for diagnosis or treatment planning.
Medical professionals using Philips MR 7700 systems with specific software versions are most at risk. Patients receiving imaging from these systems may be affected if the stiffness values are misinterpreted.
Check for product codes 782120 or 782153. Systems with software versions R11.1 or R12.1 are affected. Serial numbers listed in the recall notice indicate specific units.
Verify the software version on your system to see if it is R11.1 or R12.1.
Software versions R11.1 and R12.1 are affected.
Check for product codes 782120 or 782153 and software versions R11.1 or R12.1.
The recall record does not specify a remedy.
We’ve drafted a message you can send Philips to request your refund or repair — edit it as you like.
Subject: Recall Z-1213-2026 — Philips Philips Hello, I own a Philips that is covered by recall Z-1213-2026 from Philips. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.