Hemochron Jr. APTT Test Cuvette Recall: Risk of Inaccurate Blood Clotting Results

Accriva Diagnostics recalls Hemochron Jr. APTT Test Cuvettes with specific lot numbers due to potential inaccurate blood clotting test results, which could lead to improper anticoagulant therapy.

🇺🇸 FDA recall #Z-0317-2027Medical deviceAccriva Diagnostics ›
✎ GovInfoNow plain-language explainer

Accriva Diagnostics is recalling Hemochron Jr. APTT Test Cuvettes that may give inaccurate blood clotting results. This could lead to improper anticoagulant therapy, increasing the risk of blood clots. Read more at govinfonow.com/recall/2255730-accriva-diagnostics-inc-recalls-hemochron-jr-whole-blood-microcoagulation-system.

Source Summary by GovInfoNow · govinfonow.com/recall/2255730-accriva-diagnostics-inc-recalls-hemochron-jr-whole-blood-microcoagulation-system · Updated November 17, 2026 · FDA enforcement record: FDA #Z-0317-2027

Cite this page

“Hemochron Jr. APTT Test Cuvette Recall: Risk of Inaccurate Blood Clotting Results.” GovInfoNow, updated November 17, 2026. https://govinfonow.com/recall/2255730-accriva-diagnostics-inc-recalls-hemochron-jr-whole-blood-microcoagulation-system. Based on FDA recall #Z-0317-2027 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0317-2027%22&limit=1).

The FDA announced on Sept. 14, 2026, that Accriva Diagnostics is recalling Hemochron Jr. APTT Test Cuvettes due to a mispositioned or absent clot activating reagent. This defect may cause erroneous high results, potentially leading to inadequate anticoagulant therapy and thrombosis.

How to tell if you have it

  • Brand: Accriva Diagnostics
  • Product: Hemochron Jr., Whole Blood Microcoagulation System, Activated Partial Thromboplastin Time (APTT) Test Cuvette
  • REF: J103
  • UDI-DI: 10711234103012
  • Lot codes: B6JCA005, C6JCA008, D6JCA011, K5JCA041, C6JCA007
  • Expiration dates: 02/28/2027, 03/31/2027, 04/30/2027, 10/31/2026
  • Distribution: Worldwide; US states of MA, MO, AK, WV, NJ, TX, VA, IL, NY; countries of China, Colombia

The hazard

The cuvettes may have a mispositioned or absent clot activating reagent, which can result in out-of-range high error codes or erroneously high results. These errors may lead to retesting or an overestimation of the patient's anticoagulation, resulting in inadequate anticoagulant therapy that could cause thrombosis.

Context

GovInfoNow's records show Accriva Diagnostics has issued six other recalls since March 30, 2022, including a directCHECK ACT-LR blood control recall on April 30, 2026, for incorrect labeling and an Act-LR Cuvette recall on Nov. 5, 2025, for sharp object risks. Other recent recalls of similar products include the Diagnostica Stago Asserachrom HPIA Kit Recall (Dec. 15, 2025), the HemosIL SynthAFax Partial Thromboplastin Time Tests Recall (Nov. 18, 2025), and the Roche Elecsys Anti-TSHR Assay Recall (Nov. 21, 2025).

How to tell if you have this one

Check for lot numbers B6JCA005 (expiring 02/2027), C6JCA008 (expiring 03/2027), D6JCA011 (expiring 04/2027), K5JCA041 (expiring 10/2026), or C6JCA007 (expiring 03/2027) on the cuvette packaging.

  • Brand
    Accriva Diagnostics
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Cuvette Lot Number

    Look for lot numbers B6JCA005 (expiring 02/2027), C6JCA008 (expiring 03/2027), D6JCA011 (expiring 04/2027), K5JCA041 (expiring 10/2026), or C6JCA007 (expiring 03/2027) on the cuvette packaging.

Where it went and when

Where it was distributed

Distributed nationwide and abroad; the notice names 9 states.

Alaska: named in the noticeAKMaine: nationwideMEVermont: nationwideVTNew Hampshire: nationwideNHWashington: nationwideWAIdaho: nationwideIDMontana: nationwideMTNorth Dakota: nationwideNDMinnesota: nationwideMNIllinois: named in the noticeILWisconsin: nationwideWIMichigan: nationwideMINew York: named in the noticeNYRhode Island: nationwideRIMassachusetts: named in the noticeMAOregon: nationwideORNevada: nationwideNVWyoming: nationwideWYSouth Dakota: nationwideSDIowa: nationwideIAIndiana: nationwideINOhio: nationwideOHPennsylvania: nationwidePANew Jersey: named in the noticeNJConnecticut: nationwideCTCalifornia: nationwideCAUtah: nationwideUTColorado: nationwideCONebraska: nationwideNEMissouri: named in the noticeMOKentucky: nationwideKYWest Virginia: named in the noticeWVVirginia: named in the noticeVAMaryland: nationwideMDDelaware: nationwideDEArizona: nationwideAZNew Mexico: nationwideNMKansas: nationwideKSArkansas: nationwideARTennessee: nationwideTNNorth Carolina: nationwideNCSouth Carolina: nationwideSCDistrict of Columbia: nationwideDCOklahoma: nationwideOKLouisiana: nationwideLAMississippi: nationwideMSAlabama: nationwideALGeorgia: nationwideGAHawaii: nationwideHITexas: named in the noticeTXFlorida: nationwideFL

FDA distribution: “Worldwide - US Nationwide distribution in the states of MA, MO, AK, WV, NJ, TX, VA, IL, NY and the countries of China, Colombia.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Named in the notice: Alaska, Illinois, Massachusetts, Missouri, New Jersey, New York, Texas, Virginia, West Virginia.

Recall timeline

FDA published this recall 58 days after the firm started it.

1. Firm started the recall: Sep 14, 202612. FDA classified it Class II: Oct 30, 202623. Published in FDA’s Enforcement Report: Nov 11, 202634. Added to GovInfoNow: Nov 17, 202642026
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. Added to GovInfoNow

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0317-2027 (September 14, 2026) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0317-2027 to see it in context.
  • Plain-language summary written by qwen3:4b on November 17, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA