Use cautionMedical device recall · FDA · Published September 14, 2026
Hemochron Jr. APTT Test Cuvette Recall: Risk of Inaccurate Blood Clotting Results
Accriva Diagnostics recalls Hemochron Jr. APTT Test Cuvettes with specific lot numbers due to potential inaccurate blood clotting test results, which could lead to improper anticoagulant therapy.
Microcoagulation test used to evaluate low dose heparin anticoagulation has a mispositioned or absent clot activating reagent, which may result in out-of-range high error code, or erroneously high results, which may lead retest, or overestimation of patient's anticoagulation, leading to inadequate anticoagulant therapy, which could result in thrombosis.
“Hemochron Jr. APTT Test Cuvette Recall: Risk of Inaccurate Blood Clotting Results.” GovInfoNow, updated November 17, 2026. https://govinfonow.com/recall/2255730-accriva-diagnostics-inc-recalls-hemochron-jr-whole-blood-microcoagulation-system. Based on FDA recall #Z-0317-2027 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0317-2027%22&limit=1).
The FDA announced on Sept. 14, 2026, that Accriva Diagnostics is recalling Hemochron Jr. APTT Test Cuvettes due to a mispositioned or absent clot activating reagent. This defect may cause erroneous high results, potentially leading to inadequate anticoagulant therapy and thrombosis.
How to tell if you have it
Brand: Accriva Diagnostics
Product: Hemochron Jr., Whole Blood Microcoagulation System, Activated Partial Thromboplastin Time (APTT) Test Cuvette
REF: J103
UDI-DI: 10711234103012
Lot codes: B6JCA005, C6JCA008, D6JCA011, K5JCA041, C6JCA007
Distribution: Worldwide; US states of MA, MO, AK, WV, NJ, TX, VA, IL, NY; countries of China, Colombia
The hazard
The cuvettes may have a mispositioned or absent clot activating reagent, which can result in out-of-range high error codes or erroneously high results. These errors may lead to retesting or an overestimation of the patient's anticoagulation, resulting in inadequate anticoagulant therapy that could cause thrombosis.
Context
GovInfoNow's records show Accriva Diagnostics has issued six other recalls since March 30, 2022, including a directCHECK ACT-LR blood control recall on April 30, 2026, for incorrect labeling and an Act-LR Cuvette recall on Nov. 5, 2025, for sharp object risks. Other recent recalls of similar products include the Diagnostica Stago Asserachrom HPIA Kit Recall (Dec. 15, 2025), the HemosIL SynthAFax Partial Thromboplastin Time Tests Recall (Nov. 18, 2025), and the Roche Elecsys Anti-TSHR Assay Recall (Nov. 21, 2025).
How to tell if you have this one
Check for lot numbers B6JCA005 (expiring 02/2027), C6JCA008 (expiring 03/2027), D6JCA011 (expiring 04/2027), K5JCA041 (expiring 10/2026), or C6JCA007 (expiring 03/2027) on the cuvette packaging.
Brand
Accriva Diagnostics
Category
Medical device
Based on the official FDA remedy
What to do — step by step
Check Cuvette Lot Number
Look for lot numbers B6JCA005 (expiring 02/2027), C6JCA008 (expiring 03/2027), D6JCA011 (expiring 04/2027), K5JCA041 (expiring 10/2026), or C6JCA007 (expiring 03/2027) on the cuvette packaging.
Where it went and when
Where it was distributed
Distributed nationwide and abroad; the notice names 9 states.
Named in the notice
Covered by “nationwide”
Not listed
FDA distribution: “Worldwide - US Nationwide distribution in the states of MA, MO, AK, WV, NJ, TX, VA, IL, NY and the countries of China, Colombia.” States read from FDA’s own wording; “not listed” means the notice does not name it.
Named in the notice: Alaska, Illinois, Massachusetts, Missouri, New Jersey, New York, Texas, Virginia, West Virginia.
Recall timeline
FDA published this recall 58 days after the firm started it.
1Firm started the recall
2FDA classified it Class II
3Published in FDA’s Enforcement Report
4Added to GovInfoNow
Dates from the agency record only; steps without a date are left out.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.
Official source: FDA Recall #Z-0317-2027 (September 14, 2026) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0317-2027 to see it in context.
Plain-language summary written by qwen3:4b on November 17, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.