Baxter TruSystem 7500 Hybrid Plus Operating Table May Drop Unexpectedly During Surgery

Baxter Healthcare recalls 1 unit of TruSystem 7500 Hybrid Plus operating tables with model 1854087 due to risk of sudden table drop during surgery, potentially causing serious injury. No patient injuries reported in three complaints.

🇺🇸 FDA recall #Z-0354-2027Medical deviceBaxter Healthcare ›
✎ GovInfoNow plain-language explainer

Baxter Healthcare is recalling one unit of their TruSystem 7500 Hybrid Plus operating table column model 1854087 because the main lift mechanism may unexpectedly drop to the lowest position during use. This could cause the tabletop to become unraised, potentially displacing a patient during surgery and causing serious injury. Read more at govinfonow.com/recall/2255614-baxter-healthcare-corporation-recalls-trusystem-7500-hybrid-plus-fc-fixed.

Source Summary by GovInfoNow · govinfonow.com/recall/2255614-baxter-healthcare-corporation-recalls-trusystem-7500-hybrid-plus-fc-fixed · Published November 17, 2026 · Updated November 17, 2026 · FDA enforcement record: FDA #Z-0354-2027

Cite this page

“Baxter TruSystem 7500 Hybrid Plus Operating Table May Drop Unexpectedly During Surgery.” GovInfoNow, updated November 17, 2026. https://govinfonow.com/recall/2255614-baxter-healthcare-corporation-recalls-trusystem-7500-hybrid-plus-fc-fixed. Based on FDA recall #Z-0354-2027 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0354-2027%22&limit=1).

Baxter Healthcare Corporation voluntarily recalled one TruSystem 7500 Hybrid Plus (FC) operating table on Sept. 29, 2026, due to a main lift mechanism hazard, according to the FDA. The device may unexpectedly drop during use, potentially causing patient injury.

How to tell if you have it

  • Brand: Baxter Healthcare
  • Product: TruSystem 7500 Hybrid Plus (FC), Fixed-Installation Operating Table Column
  • Model/Catalog Number: 1854087
  • Serial numbers: 107971925
  • UDI-DI number: 00887761974067

The hazard

The main lift mechanism may unexpectedly drop from its set position to its lowest position during use, and the tabletop could not be raised again after the drop. Baxter has received three complaints related to this issue, with no reported patient injury.

Context

GovInfoNow's records show Baxter Healthcare Corporation has 190 other recalls on record since Sept. 2, 2015. Other recent recalls of similar products include the IMRIS ORT300 Operating Room Table Recall (Aug. 6, 2026), the IMRIS ORT200 Operating Room Table Recall (Aug. 6, 2026), and the Philips Azurion 7 M12 System Recall (Jan. 26, 2026).

How to tell if you have this one

Check for the model catalog number 1854087 on the TruSystem 7500 Hybrid Plus operating table column. The serial numbers 107971925 and UDI-DI number 00887761974067 are also associated with this recall.

  • Brand
    Baxter Healthcare
  • Category
    Medical device
FDA recall event · 5 products

All 5 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class II
  • Status
    Ongoing
  • Distribution
    Worldwide - US Nationwide distribution.
Recall numberProduct, codes & lotsQuantity
Z-0350-2027TruSystem 7500 surgical table, Model/Catalog Number: 4091000; Serial numbers: 108491706, 108726496, 108758657, 108858029, 108945825, 109083212, 109343910, 109684319, 110016886, 108557062, 108736192, 108773032, 108882327, 108976094, 109174333, 109370104, 109690845, 110016887, 108574826, 108736193, 108773033, 108882328, 108976095, 109174334, 109396256, 109690846, 110072583, 108578755, 108736194, 108842519, 108883564, 108976101, 109174335, 109464647, 109708813, 110084982, 108580717, 108736195, 108842523, 108883565, 109047445, 109174336, 109464648, 109708816, 110086138, 108580718, 108743656, 108842524, 108896687, 109047446, 109228237, 109464649, 109885369, 110144223, 108580720, 108743657, 108842525, 108896688, 109047447, 109228238, 109464650, 109896735, 110144224, 108604827, 108745163, 108842526, 108912913, 109047448, 109249973, 109488031, 109896736, 110144225, 108604828, 108745164, 108842679, 108912914, 109047451, 109249974, 109493243, 109896737, 110158878, 108604832, 108745165, 108842680, 108931731, 109047452, 109249989, 109500155, 109896738, 110158879, 108604833, 108745166, 108849690, 108931732, 109047460, 109301944, 109561673, 109933665, 110175899, 108726492, 108745180, 108849691, 108945772, 109047461, 109301945, 109561674, 109948691, 110175900, 108726493, 108745181, 108849694, 108945823, 109078531, 109301946, 109571317, 110007863, 110163294, 108726495, 108758656, 108849695, 108945824, 109078533, 109343482, 109571318, 110016883, 110163293, 110183861, 110220396, 110220395; UDI-DI number: 00887761968639129 units
Z-0351-2027Operating table column TS7500 MOBIUS, Hybrid operating table column, Model/Catalog Number: 1704695; Serial numbers: 108094281, 108468075, 108484417, 108506318, 108516636, 108536785, 108544115, 108604154, 108463223, 108484404, 108489634, 108511411, 108516637, 108544106, 108544116, 108618964, 108463224, 108484405, 108489648, 108511412, 108521436, 108544107, 108604137, 108618965, 108468062, 108484406, 108506316, 108511413, 108521452, 108544110, 108604140, 108468063, 108484408, 108506317, 108516634, 108536776, 108544111, 108604153; UDI-DI number: 0088776197442577 units
Z-0352-2027Operating Table TruSystem 7000 U, Surgical table, universal, Model/Catalog Number: 1604788; Serial numbers: 108517025, 110115554, 110144247, 110188916, 109558596, 110139748, 110166201, 110188918, 109737582, 110139769, 110177678, 110192686, 110115551, 110144245, 110185616, 109889060; UDI-DI number: 0088776196752616 units
Z-0353-2027Operating Table TruSystem 7000 U (dV), Model/Catalog Number: 1723633; Serial numbers: 108920104, 110192607, 110192885, 110208936, 110212885, 110220066, 110227520, 110248337, 110250614, 109423464, 110192608, 110192886, 110208941, 110212886, 110220068, 110227523, 110248343, 110250621, 109517138, 110192609, 110206656, 110210717, 110220055, 110220069, 110227524, 110249706, 110250622, 110139767, 110192683, 110206657, 110210718, 110220056, 110220071, 110227525, 110249712, 110206662, 110139776, 110192882, 110206659, 110210720, 110220057, 110220072, 110227552, 110250605, 110188921, 110188919, 110192883, 110206660, 110210721, 110220058, 110220078, 110227553, 110250609, 110188922, 110188920, 110192884, 110206664, 110212881, 110220065, 110220079, 110248218, 110250613; UDI-DI number: 0088776196751962 units
Z-0354-2027TruSystem 7500 Hybrid Plus (FC), Fixed-Installation Operating Table Column, Model/Catalog Number: 1854087; Serial numbers: 107971925; UDI-DI number: 008877619740671 unit
Based on the official FDA remedy

What to do — step by step

  1. Stop using it now

    Immediately cease using the TruSystem 7500 Hybrid Plus operating table column model 1854087 with serial numbers 107971925 or UDI-DI number 00887761974067.

Where it went and when

Where it was distributed

Distributed nationwide, and outside the U.S.

Alaska: yesAKMaine: yesMEVermont: yesVTNew Hampshire: yesNHWashington: yesWAIdaho: yesIDMontana: yesMTNorth Dakota: yesNDMinnesota: yesMNIllinois: yesILWisconsin: yesWIMichigan: yesMINew York: yesNYRhode Island: yesRIMassachusetts: yesMAOregon: yesORNevada: yesNVWyoming: yesWYSouth Dakota: yesSDIowa: yesIAIndiana: yesINOhio: yesOHPennsylvania: yesPANew Jersey: yesNJConnecticut: yesCTCalifornia: yesCAUtah: yesUTColorado: yesCONebraska: yesNEMissouri: yesMOKentucky: yesKYWest Virginia: yesWVVirginia: yesVAMaryland: yesMDDelaware: yesDEArizona: yesAZNew Mexico: yesNMKansas: yesKSArkansas: yesARTennessee: yesTNNorth Carolina: yesNCSouth Carolina: yesSCDistrict of Columbia: yesDCOklahoma: yesOKLouisiana: yesLAMississippi: yesMSAlabama: yesALGeorgia: yesGAHawaii: yesHITexas: yesTXFlorida: yesFL

FDA distribution: “Worldwide - US Nationwide distribution.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Distributed nationwide, and outside the U.S.

Recall timeline

FDA published this recall 43 days after the firm started it.

1. Firm started the recall: Sep 29, 202612. FDA classified it Class II: Nov 3, 202623. Published in FDA’s Enforcement Report: Nov 11, 202634. Added to GovInfoNow: Nov 17, 202642026
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. Added to GovInfoNow

Dates from the agency record only; steps without a date are left out.

Similar recalls

Other Baxter Healthcare recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0354-2027 (September 29, 2026) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0354-2027 to see it in context.
  • Plain-language summary written by qwen3:4b on November 17, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA