The FDA announced on October 1, 2026, that restor3d Inc. is recalling Veritas Reverse Total Shoulder Replacement implants due to a hazard of potential dislocation or reduced range of motion. The recall affects units distributed nationwide and in specific states.
How to tell if you have it
- Brand: restor3d
- Product: Veritas Reverse Total Shoulder Replacement
- Model/Catalog Numbers: 3500-140C3, 3500-140U3, 3500-140C1, 3500-140U1
- Sizes: 1 and 2 with 40mm Polymer Liner
- Lot codes/UDIs: Product Code 3500-140C3 UDI/GTIN 00840097513492; Product Code 3500-140U3 UDI/GTIN 00840097513461; Product Code 3500-140C1 UDI/GTIN 00840097513430; Product Code 3500-140U1 UDI/GTIN 00840097513409
- Distribution: US Nationwide, Texas, Massachusetts, Oklahoma, Florida, Colorado, North Carolina
- Quantity: 240 units
The hazard
According to the FDA, pairing shoulder implant stem sizes 1 and 2 with the largest glenosphere diameter (40 mm) creates excessive mechanical loading. This configuration may result in polyethylene components lacking adequate retention strength, potentially causing dislocation, reduced range of motion, or revision surgery.
Context
GovInfoNow's records show restor3d Inc. has issued four other recalls since March 30, 2022. Recent restor3d recalls include a September 4, 2026, notice for non-sterilized MTP Implants and a July 27, 2026, recall of Kinos Axiom Ankle Implants due to incorrect cutting guides. Other recent recalls of similar products include the Zimmer Off-Axis Shoulder System recall on May 20, 2026, and the Diversatek Viper 3-Stage Balloon Dilator recall on March 18, 2026.