Lower riskMedical device recall · FDA · Published September 11, 2026
BioMerieux EMAG System May Cause Incorrect Test Results Due to Software Error
BioMerieux recalls EMAG nucleic acid sample preparation instruments with model 418591/418591U that may show delayed or no results due to a software error after firmware updates. Affected units include specific serial numbers sold worldwide.
Delayed or no result when analyzing clinical samples with EMAG system (ref: 418591/418591U) due to a randomly occurring software error, Error 10003 Heater timeout after updating to recent firmware SP4 Patch 2
“BioMerieux EMAG System May Cause Incorrect Test Results Due to Software Error.” GovInfoNow, updated November 11, 2026. https://govinfonow.com/recall/2090905-biomerieux-sa-recalls-emag-system-model-catalog-number-418591-418591u. Based on FDA recall #Z-0301-2027 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0301-2027%22&limit=1).
The FDA announced on Sept. 11, 2026, that BioMerieux SA is recalling the EMAG Nucleic Acid Sample Prep System due to a software error that can cause delayed or no results when analyzing clinical samples. The agency issued recall number Z-0301-2027 for the voluntary Class II recall.
The recall addresses a randomly occurring software error, specifically Error 10003 Heater timeout, which occurs after updating to recent firmware SP4 Patch 2. This issue may result in delayed or no results when analyzing clinical samples with the EMAG system.
Context
GovInfoNow's records show BioMerieux SA has issued six other recalls since Oct. 18, 2017, including a Sept. 2, 2021, recall for MYLA Software Versions 4.7-4.8.2 due to false negative risk and a June 22, 2021, recall for MYLA Software due to unwanted result alterations. Other recent recalls of similar products include the June 4, 2026, Biomerieux VITEK 2 AST-N447 Kit Recall for false resistance results risk and the July 24, 2026, Biomerieux VITEK 2 AST-N461 Test Kit Recall for false resistant results risk.
How to tell if you have this one
Check for model number 418591/418591U on the device. Affected units have serial numbers listed in the recall (e.g., IM03343, IM03283). The system was sold worldwide starting in 2026.
Brand
BioMerieux
Category
Medical device
Based on the official FDA remedy
What to do — step by step
Check for Error 10003
Look for the heater timeout error message after firmware updates to SP4 Patch 2.
Contact BioMerieux
Reach out to BioMerieux for specific guidance on troubleshooting or replacement options.
Where it went and when
Where it was distributed
Distributed nationwide and abroad; the notice names 9 states.
Named in the notice
Covered by “nationwide”
Not listed
FDA distribution: “Worldwide - US Nationwide distribution in the states of CA, IN, KS, LA, MN, MO, NY, OH, and TX. The countries of Canada, United Kingdom, Sweden, France, Norway, Netherlands, Belgium, Denmark, Germany, Austria.” States read from FDA’s own wording; “not listed” means the notice does not name it.
Named in the notice: California, Indiana, Kansas, Louisiana, Minnesota, Missouri, New York, Ohio, Texas.
Recall timeline
FDA published this recall 54 days after the firm started it.
1Firm started the recall
2FDA classified it Class II
3Published in FDA’s Enforcement Report
4Added to GovInfoNow
Dates from the agency record only; steps without a date are left out.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.
Official source: FDA Recall #Z-0301-2027 (September 11, 2026) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0301-2027 to see it in context.
Plain-language summary written by qwen3:4b on November 11, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.