Beckman Coulter recalls 92 Optima XPN ultracentrifuge rotors with reference number 331302 that may break during use, risking sample destruction and lab exposure to biohazardous materials.
Optima XE XPN ultracentrifuge rotors may contain a microstructure defect that can cause rotors to break into two or more pieces (bifurcate) during a centrifugation run, resulting in destruction of samples being processed. When used for in vitro diagnostic (IVD) sample preparation, rotor breakage may delay patient test results. Additionally, sample destruction may expose laboratory personnel to biohazardous materials.
“Beckman Coulter Optima XPN Ultracentrifuge Rotors Recall: Microstructure Defect Risk.” GovInfoNow, updated November 11, 2026. https://govinfonow.com/recall/1932664-beckman-coulter-inc-recalls-rotor-assy-sw40-reference-number-331302. Based on FDA recall #Z-0201-2027 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0201-2027%22&limit=1).
The FDA announced on Sept. 21, 2026, that Beckman Coulter, Inc. is recalling Optima XPN Ultracentrifuge Rotors due to a microstructure defect that may cause rotors to break during use. The agency said the hazard can result in sample destruction and potential exposure of laboratory personnel to biohazardous materials.
How to tell if you have it
Brand: Beckman Coulter
Product: Optima XPN Ultracentrifuge Rotors (ROTOR ASSY-SW40, Reference Number 331302)
Compatible Centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima XPN-90 (Model Numbers A99842 and A94468), Optima XPN-80 (Model Numbers A99839 and A95765), Optima XE-100 (Model Numbers A99836 and A94516), and Optima XE-90 (Model Numbers A99833 and A94471)
According to the FDA, the Optima XE XPN ultracentrifuge rotors may contain a microstructure defect that can cause them to bifurcate during a centrifugation run. This breakage can destroy samples being processed and, when used for in vitro diagnostic sample preparation, may delay patient test results. Additionally, sample destruction may expose laboratory personnel to biohazardous materials.
Context
GovInfoNow's records show Beckman Coulter, Inc. has 135 other recalls on record since 2017-06-20. Other recent recalls of similar products include DxC 500 AU Clinical Chemistry Analyzers (2026-09-02), DxU 840m Iris Recall (2026-08-12), and DxC 500i Modules with Software 1.5 Risk (2026-07-15). Other recent recalls of similar products include Philips IQon Spectral CT Recall (2026-01-07), Philips Spectral CT Recall (2026-01-07), and Philips Brilliance iCT CT Scanner Recall (2026-01-07).
How to tell if you have this one
Check for rotors with reference number 331302 installed in Optima XPN-100 centrifuges (model numbers A99846 or A94469). Serial numbers range from 24U15260 to 26U15764. The Optima UDI-DI codes match specific model numbers with serial ranges.
Brand
Beckman Coulter
Category
Medical device
FDA recall event · 2 products
All 2 products in this recall
FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.
Classification
Class II
Status
Ongoing
Distribution
US Nationwide distribution.
Recall number
Product, codes & lots
Quantity
Z-0201-2027
ROTOR ASSY-SW40, Reference Number 331302, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima XPN-90 (Model Numbers A99842 and A94468), Optima XPN-80 (Model Numbers A99839 and A95765), Optima XE-100 (Model Numbers A99836 and A94516), and Optima XE-90 (Model Numbers A99833 and A94471); Serial Numbers: 24U15260 to 26U15764; Optima UDI-DI: A99846 - 15099590416324, A99842 - 15099590415754, A99839 - 15099590415822, A99836 - 15099590415846, A99833 - 15099590415860
92 units
Z-0202-2027
ROTOR ASSY-SW41, Reference Number 331362, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima XPN-90 (Model Numbers A99842 and A94468), Optima XPN-80 (Model Numbers A99839 and A95765), Optima XE-100 (Model Numbers A99836 and A94516), and Optima XE-90 (Model Numbers A99833 and A94471); Serial Numbers: 24U15250 to 26U15766; Optima UDI-DI: A99846 - 15099590416324, A99842 - 15099590415754, A99839 - 15099590415822, A99836 - 15099590415846, A99833 - 15099590415860
336 units
Based on the official FDA remedy
What to do — step by step
Stop using affected rotors
Immediately remove rotors with reference number 331302 installed in Optima XPN-100 centrifuges (model numbers A99846 or A94469) if they have serial numbers from 24U15260 to 26U15764.
Where it went and when
Where it was distributed
Distributed nationwide.
Nationwide
Not listed
FDA distribution: “US Nationwide distribution.” States read from FDA’s own wording; “not listed” means the notice does not name it.
Distributed nationwide.
Recall timeline
FDA published this recall 37 days after the firm started it.
1Firm started the recall
2FDA classified it Class II
3Published in FDA’s Enforcement Report
4Added to GovInfoNow
Dates from the agency record only; steps without a date are left out.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.
Official source: FDA Recall #Z-0201-2027 (September 21, 2026) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0201-2027 to see it in context.
Plain-language summary written by qwen3:4b on November 11, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.