Use cautionMedical device recall · FDA · Published September 24, 2026
Philips Lumify Ultrasound System May Disconnect During Scans
Philips recalls Lumify Diagnostic Ultrasound Systems with specific Android and iOS software versions due to potential disconnections during scans, causing diagnosis delays.
Ultrasound system may: 1) Not recognize previous transducer registration, prevent scanning, require re-registration with Wi-Fi/cellular access, which may be unavailable outside of the hospital (ambulances/aircraft/field); 2) At connection between transducer USB cable to tablet/phone, there may be an unstable connection/may disconnect disrupting function. May cause diagnosis/treatment delay.
“Philips Lumify Ultrasound System May Disconnect During Scans.” GovInfoNow, updated November 11, 2026. https://govinfonow.com/recall/1932660-philips-ultrasound-llc-recalls-philips-lumify-diagnostic-ultrasound-system. Based on FDA recall #Z-0152-2027 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0152-2027%22&limit=1).
Philips Ultrasound, LLC recalled the Lumify Diagnostic Ultrasound System on Sept. 24, 2026, according to the FDA. The device may fail to recognize transducer registration or disconnect during use, potentially delaying diagnosis or treatment.
The ultrasound system may not recognize previous transducer registration, preventing scanning and requiring re-registration with Wi-Fi or cellular access that may be unavailable in ambulances, aircraft, or the field. An unstable connection at the transducer USB cable interface may also disconnect, disrupting function and causing diagnosis or treatment delays.
Context
GovInfoNow's records show Philips Ultrasound, LLC has issued three other recalls since July 14, 2022. On Aug. 7, 2026, the company recalled the EPIQ Elite Ultrasound System due to inaccurate blood flow measurements. Other recent recalls of similar products include the Philips Lumify Diagnostic Ultrasound System (Oct. 31, 2025) and the Philips EPIQ Ultrasound Systems with X5-1c Transducer (July 14, 2022).
How to tell if you have this one
Check for Android software versions 00884838122116/1.X, 3.X, 4.X, 5.X or iOS versions 00884838124141/2.X and 5.X on the system.
Brand
Philips
Category
Medical device
Based on the official FDA remedy
What to do — step by step
Check Software Version
Verify the Android or iOS software version on your Lumify system to confirm if it is in the recalled versions.
Where it went and when
Where it was distributed
Distributed nationwide and abroad; the notice names 50 states.
Named in the notice
Covered by “nationwide”
Not listed
FDA distribution: “Worldwide - US Nationwide distribution including in the states of CA, MA, IL, NY, OR, GA, TX, WA, FL, PA, WI, LA, NJ, KY, VA, MS, AZ, SC, CT, MN, UT, OH, AL, MI, NH, MO, TN, IN, DC, NE, NM, MT, NV, NC, MD, IA, SD, VT, KS, AK, CO, ID, OK,…” States read from FDA’s own wording; “not listed” means the notice does not name it.
Named in the notice: Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West Virginia, Wisconsin, Wyoming.
Recall timeline
FDA published this recall 34 days after the firm started it.
1Firm started the recall
2FDA classified it Class I
3Published in FDA’s Enforcement Report
4Added to GovInfoNow
Dates from the agency record only; steps without a date are left out.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.
Official source: FDA Recall #Z-0152-2027 (September 24, 2026) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0152-2027 to see it in context.
Plain-language summary written by qwen3:4b on November 11, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.