Skip to content
Use caution Medical device recall · FDA · Published January 2, 2018

Stryker Reprocessed Catheters Recall: EEPROM Error with CARTO EP Navigation Systems

Stryker recalls reprocessed BW Lasso 2515 NAV eco Variable Diagnostic EP Catheters due to potential EEPROM chip errors when used with CARTO EP Navigation Systems version 2.3 or higher.

Take action

Contact the company

Recalling firmStryker Sustainability Solutions1810 W Drake Dr · Tempe, AZ 85283-4327
✎ GovInfoNow plain-language explainer

What happened, in plain English

Stryker is recalling reprocessed BW Lasso 2515 NAV eco Variable Diagnostic EP Catheters that may have EEPROM chip errors when used with CARTO EP Navigation Systems. This could lead to incorrect cardiac mapping during procedures, potentially affecting patient safety. Read more at govinfonow.com/recall/72196-stryker-reprocessed-catheters-recall-eeprom-error-with-carto-ep-navigation.

Source Summary by GovInfoNow · govinfonow.com/recall/72196-stryker-reprocessed-catheters-recall-eeprom-error-with-carto-ep-navigation · Updated August 31, 2026 · FDA enforcement record: FDA #Z-0916-2018

Cite this page

“Stryker Reprocessed Catheters Recall: EEPROM Error with CARTO EP Navigation Systems.” GovInfoNow, updated August 31, 2026. https://govinfonow.com/recall/72196-stryker-reprocessed-catheters-recall-eeprom-error-with-carto-ep-navigation. Based on FDA recall #Z-0916-2018 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0916-2018%22&limit=1).

Why it’s dangerous

The catheters may display incorrect error codes when used with CARTO EP Navigation Systems version 2.3 or higher, which could lead to improper cardiac mapping during electrophysiology procedures. This might result in ineffective treatment or unintended complications.

Who’s affected

Healthcare providers using these catheters for cardiac electrophysiology procedures are most at risk. Patients undergoing such procedures with the recalled catheters could experience inaccurate mapping results.

How to tell if you have this one

Check for the serial numbers listed in the recall notice (e.g., 2451387, 2484264, etc.). The catheters are reprocessed BW Lasso 2515 NAV eco Variable Diagnostic EP Catheters with reference number D134301.

  • Brand
    Stryker Sustainability Solutions
  • Category
    Medical device
FDA recall event · 2 products

All 2 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class II
  • Status
    Terminated
  • Distribution
    USA (nationwide) Distribution was made to medical facilities in AZ, CA, CT, FL, GA, IL, KS, LA, MN, MO, OH, PA, TX, VT, WA, and Wi. Foreign distribution was made to Canada.
Recall numberProduct, codes & lotsQuantity
Z-0916-2018Stryker Sustainability Solutions Reprocessed BW Lasso 2515 NAV eco Variable Diagnostic EP Catheters, Ref D134301, Sterile, Rx. The Reprocessed 2515 NAV eco Variable Electrophysiology (EP) Catheters are indicated for multiple electrode electrophysiological mapping of the cardiac structures of the heart, i.e. recording or stimulation only. They are designed to obtain electrograms in the atrial regions of the heart. The Reprocessed 2515 NAV eco Variable EP Catheters provide location information when used with compatible CARTO EP Navigation Systems version 2.3 or higher.; Serial numbers 2451387, 2484264, 2484471, 2486823, 2487380, 2488350, 2490592, 2490605, 2492930, 2494064, 2494067, 2495242, 2495335, 2495337, 2495339, 2500403, 2500911, 2510524, 2518512, 2523848, 2523849, 2523851, 2525678, 2528071, 2534854, 2534856, 2608429, 2609031, 2609102, 2609105, 2609106, 2609107, 2609985, 2610794, 2610802, 2614613, 2614847, 2614909, 2615094, 2615098, 2615099, 2615319, 2615646, 2615712, 2615722, 2615725, 2615937, 2615938, 2615940, 2615943, 2615952, 2615966, 2616087, 2616088, 2616089, 2616090, 2616212, 2617794, 2617920, 2617987, 2617999, 2618010, 2618258, 2618832, 2618837, 2618881, 2618884, 2618990, 2618992, 2618993, 2619615, 2619742, 2619744, 2619827, 2620937, 2620942, 2620945, 2621283, 2621384, 2621385, 2621387, 2621424, 2621425, 2621847, 2621860, 2621880, 2623972, 2623974, 2624006, 2624010, 2624113, 2624138, 2625427, 2625536, 2625550, 2625623, 2625625, 2626330, 2626348, 2626929, 2626966, 2626988, 2627596, 2627946, 2627948, 2627949, 2629162, 2629164, 2630084, 2630085, 2630087, 2630351, 2631186, 2631188, 2631189, 2631191, 2631680, 2632641, 2633968, 2634008, 2634252, 2634271, 2634596, 2634709, 2634815, 2634836, 2635075, 2635077, 2635342, 2635343, 2635867, 2635873, 2636658, 2636759, 2636906, 2636986, 2637009, 2637022, 2637023, 2637233, 2637234, 2637237, 2637239, 2637257, 2637308, 2637495, 2637497, 2640103, 2640106, 2640107, 2640118, 2640119, 2642853, 2643222, 2643426, 2643432, 2646002, 2647122, 2669947, 2693960, 2696255, 2705129, 2708915, 2708916, 2712736, 2712756, 2713556, 2713593, 2713595, and 2713596.170 devices
Z-0917-2018Stryker Sustainability Solutions Reprocessed BW Lasso 2515 NAV eco Variable Diagnostic EP Catheters, Ref D134302, Sterile, Rx. The Reprocessed 2515 NAV eco Variable Electrophysiology (EP) Catheters are indicated for multiple electrode electrophysiological mapping of the cardiac structures of the heart, i.e. recording or stimulation only. They are designed to obtain electrograms in the atrial regions of the heart. The Reprocessed 2515 NAV eco Variable EP Catheters provide location information when used with compatible CARTO EP Navigation Systems version 2.3 or higher.; Serial numbers 2468846, 2468849, 2509029, 2489074, 2491174, 2500529, 2465989, and 25005288 devices
Based on the official FDA remedy

What to do — step by step

  1. Check Serial Numbers

    Verify if your catheters have any of the listed serial numbers (e.g., 2451387, 2484264, etc.).

Where it went and when

Where it was distributed

Distributed nationwide and abroad; the notice names 16 states.

Alaska: nationwideAKMaine: nationwideMEVermont: named in the noticeVTNew Hampshire: nationwideNHWashington: named in the noticeWAIdaho: nationwideIDMontana: nationwideMTNorth Dakota: nationwideNDMinnesota: named in the noticeMNIllinois: named in the noticeILWisconsin: named in the noticeWIMichigan: nationwideMINew York: nationwideNYRhode Island: nationwideRIMassachusetts: nationwideMAOregon: nationwideORNevada: nationwideNVWyoming: nationwideWYSouth Dakota: nationwideSDIowa: nationwideIAIndiana: nationwideINOhio: named in the noticeOHPennsylvania: named in the noticePANew Jersey: nationwideNJConnecticut: named in the noticeCTCalifornia: named in the noticeCAUtah: nationwideUTColorado: nationwideCONebraska: nationwideNEMissouri: named in the noticeMOKentucky: nationwideKYWest Virginia: nationwideWVVirginia: nationwideVAMaryland: nationwideMDDelaware: nationwideDEArizona: named in the noticeAZNew Mexico: nationwideNMKansas: named in the noticeKSArkansas: nationwideARTennessee: nationwideTNNorth Carolina: nationwideNCSouth Carolina: nationwideSCDistrict of Columbia: nationwideDCOklahoma: nationwideOKLouisiana: named in the noticeLAMississippi: nationwideMSAlabama: nationwideALGeorgia: named in the noticeGAHawaii: nationwideHITexas: named in the noticeTXFlorida: named in the noticeFL

FDA distribution: “USA (nationwide) Distribution was made to medical facilities in AZ, CA, CT, FL, GA, IL, KS, LA, MN, MO, OH, PA, TX, VT, WA, and Wi. Foreign distribution was made to Canada.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Named in the notice: Arizona, California, Connecticut, Florida, Georgia, Illinois, Kansas, Louisiana, Minnesota, Missouri, Ohio, Pennsylvania, Texas, Vermont, Washington, Wisconsin.

Recall timeline

FDA terminated this recall 4 months after the firm started it.

1. Firm started the recall: Jan 2, 201812. FDA classified it Class II: Mar 6, 201823. Published in FDA’s Enforcement Report: Mar 14, 201834. FDA terminated the recall: May 2, 201842018
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-0916-2018 (January 2, 2018) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-0916-2018 to see it in context.
  • Plain-language summary written by qwen3:4b on August 31, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA