The FDA announced on Jan. 4, 2018, that BioMerieux, Inc. voluntarily recalled its VIDAS System combined with Quality Control VIDAS (QCV) products due to a Class II hazard. The recall affects 1,866 units distributed worldwide and in all U.S. states.
How to tell if you have it
- VIDAS ANALYSER 110/220V, 93296 (03573026133573)
- VIDAS ANALYSER 220V, 93297 (03573026133580)
- VIDAS ANALYSEUR, 99011 (03573026109257)
- VIDAS ANALYSER EU, 99444 (03573026114060)
- VIDAS ANALYSER US, 99445 (03573026114084)
- VIDAS ANALYSER, 99735 (03573026140427)
- VIDAS ANALYSER US, W2205 (03573026029678)
- VIDAS ANALYSER, 410417 (03573026351663)
- Quality Control VIDAS (QCV), 30706 (03573026121631)
- VIDAS ANALYSER EU, W3205 (03573026030292)
- VIDAS INDUSTRY US, 421002 (03573026553319)
The hazard
According to the FDA, the recall addresses an increase in events affecting performance regarding pump clogging and displacement of the color-coded SPR label. These issues could contribute to erroneous results for any assay performed on the VIDAS and mini VIDAS systems due to a decrease of pipetted volume.
Context
GovInfoNow's records show Biomerieux Inc has had 36 other recalls on record since July 12, 2017. Recent recalls include the Biomerieux VITEK 2 AST-N461 Test Kit Recall (July 24, 2026), the Biomerieux VITEK 2 AST-N447 Kit Recall (June 4, 2026), and the Biomerieux VITEK 2 Gram-negative Susceptibility Cards recall (Jan. 5, 2026). Other recent recalls of similar products include the Bioporto ProNephro AKI Reagent Kit Recall (May 18, 2026), the Vivoo pH Test Recall (Nov. 22, 2025), and the Vivoo Vaginal pH Test Recall (Nov. 22, 2025).