The FDA announced on March 23, 2018, that Medtronic Navigation, Inc. is recalling Visualase Cooled Laser Applicator Systems due to potential low laser energy output or fiber burnout. The recall affects 1,283 devices distributed worldwide.
How to tell if you have it
- Brand: Medtronic
- Product: Visualase Cooled Laser Applicator System (VCLAS)
- Part Number 9735559, Visualase Cooled Laser Applicator System 3mm Tip
- Part Number 9735560 Visualase Cooled Laser Applicator System 10mm Tip
- Part Number 9735561 Visualase Cooled Laser Applicator System 15mm Tip
The hazard
Certain lots of the VCLAS 3MM, 10MM, and 15MM laser diffusing fibers were not tested at intended laser power levels due to a calibration issue with test equipment. This defect can lead to the potential for low laser energy output from the fiber or fiber burnout at lower wattage.
Context
GovInfoNow's records show Medtronic Navigation, Inc. has 25 other recalls on record since June 7, 2017. Recent brand recalls include issues with PIN surgical pins not fitting patient reference frames reported on Dec. 4, 2024, and software anomalies causing flipped images or missing warnings in StealthStation S8 App versions 1.3.0 and 1.3.2 reported on Aug. 22, 2024. Other recent recalls of similar products include the Lutronic DermaV Laser System recall for risk of patient burn during treatment (Aug. 6, 2026), the Diowave Laser System recall requiring a software update for safety (Jan. 9, 2026), and the Lutronic CLARITY II Laser System Model 1110200210 recall for sparks and potential patient burns (Jan. 27, 2026).