The FDA announced on April 13, 2018, that CareFusion 303, Inc. is recalling Alaris Pumps, Model 8100, with bezel mechanical assemblies made of FR-110 plastic resin due to a potential weakening of the plastic. The agency classified this as a Class I recall affecting approximately 607,618 units distributed worldwide.
How to tell if you have it
- Brand: CareFusion
- Product: Alaris Pump, Model 8100
- Component: Bezel mechanical assemblies manufactured with FR-110 plastic resin
- Manufacture Date: April 2011 through June 2017
- Affected Parts: LVP Mechanism Sub Assembly (P/N 10942012, P/N 49000007, and P/N 49000203); Bezel Kit Assembly (P/N 10964559 and P/N 49000204)
- Quantity: 607,618 units
The hazard
According to the FDA, the manufacturing process for the FR-110 plastic may have weakened the bezel, leading to separation of bezel posts or external cracking over time. This damage can result in free flow, over infusion, under infusion, or interruption of infusion.
Context
GovInfoNow's records show that CareFusion 303, Inc. has had 64 other recalls on record since June 12, 2017. Recent brand recalls include a BD Pyxis Mini Main REF 349 Medication Cabinet recall announced on April 30, 2026, due to a fingerprint scanner overheating risk, and a CareFusion recall from February 11, 2026, regarding unclear Transport Layer Security labeling. Other recent recalls of similar products include a BD Alaris Pump Infusion Sets Recall announced on August 19, 2026, for performance data discrepancies that may cause inappropriate pump function.