Skip to content
Lower risk Medical device recall · FDA · Published April 30, 2018

Polymedco PATHFAST Myoglobin II Cartridge Recall: Leakage Risk

Polymedco recalls PATHFAST Myoglobin II reagent cartridges with a pinhole causing leakage, potentially leading to inaccurate test results. Affected units may have measurement errors up to +20%.

🇺🇸 FDA recall #Z-3185-2018Medical devicePolymedco ›
Take action

Contact the company

Recalling firmPolymedco, Inc.,/Polymedco Cancer Diagnostic Products, Llc510 Furnace Dock Rd · Cortlandt Manor, NY 10567-6220
✎ GovInfoNow plain-language explainer

Polymedco is recalling PATHFAST Myoglobin II reagent cartridges that may have a pinhole causing leakage. This could lead to test results being up to 20% higher than actual values, which might affect cancer diagnosis accuracy. Read more at govinfonow.com/recall/71262-polymedco-pathfast-myoglobin-ii-cartridge-recall-leakage-risk.

Source Summary by GovInfoNow · govinfonow.com/recall/71262-polymedco-pathfast-myoglobin-ii-cartridge-recall-leakage-risk · Updated August 21, 2026 · FDA enforcement record: FDA #Z-3185-2018

Cite this page

“Polymedco PATHFAST Myoglobin II Cartridge Recall: Leakage Risk.” GovInfoNow, updated August 21, 2026. https://govinfonow.com/recall/71262-polymedco-pathfast-myoglobin-ii-cartridge-recall-leakage-risk. Based on FDA recall #Z-3185-2018 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3185-2018%22&limit=1).

The FDA announced on April 30, 2018, that Polymedco, Inc. voluntarily recalled PATHFAST Myoglobin II reagent cartridges due to a pinhole hazard that may cause leakage and inaccurate test results. The agency issued recall number Z-3185-2018 for the Class II device.

How to tell if you have it

  • Brand: Polymedco
  • Product: PATHFAST Myoglobin II reagent cartridge / PATHFAST NTproBNP
  • Catalog Number: PF1111-K
  • Lot Code: T738
  • Expiration Date: 6/30/2018
  • UPC: 04987595301230
  • Distribution: AR, CA, FL, ID, LA, MN, NY and TX

The hazard

A pinhole caused by the laser printer in the aluminum sheet over the ALP reagent wells may cause leakage through the pinhole. If a cartridge with leakage is used, higher measurement values with a bias as low as +12% and no higher than +20% may be obtained.

Context

GovInfoNow's records show Polymedco has two other recalls on record since October 6, 2025, including a August 25, 2026 recall of SEDIPLAST Diagnostic Tubes for blood exposure risk and an October 6, 2025 recall of OC-Auto FOBT-CHEK Negative Controls for incorrect expiration date labeling. Other recent recalls of similar products include Milestone Vacuum Bag SU Large Specimen Container Recall (March 9, 2026) and Siemens Dimension Tacrolimus Flex Reagent Cartridge Recall (May 7, 2026).

How to tell if you have this one

Check for the catalog number PF1111-K on the cartridge. The affected lot is T738, which expires June 30, 2018.

  • Brand
    Polymedco
  • Product
    Polymedco, Inc.,/Polymedco Cancer Diagnostic Products, Llc recalls PATHFAST Myoglobin II reagent cartridge Catalog Number: PF1111-K PATHFAST¿ NTproBNP is a…
  • Category
    Medical device
FDA recall event · 2 products

All 2 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class II
  • Status
    Terminated
  • Distribution
    US Distribution to states of: AR, CA, FL, ID, LA, MN, NY and TX.
Recall numberProduct, codes & lots
Z-3185-2018PATHFAST Myoglobin II reagent cartridge Catalog Number: PF1111-K PATHFAST¿ NTproBNP is a product for in-vitro diagnostic use with the in vitro diagnostic system PATHFAST¿ for the quantitative measurement of N-terminal-pro B-type Natriuretic Peptide (NT-proBNP) in heparinized or EDTA whole blood and plasma.; Lot T738 Exp: 6/30/2018 UDI: (01)04987595301230(17)180630(10)T738
Z-3186-2018PATHFAST NTproBNP Catalog Number: PF1061-KUS PATHFAST¿ NTproBNP is a product for in-vitro diagnostic use with the in vitro diagnostic system PATHFAST¿ for the quantitative measurement of N-terminal-pro B-type Natriuretic Peptide (NT-proBNP) in heparinized or EDTA whole blood and plasma.; Lot : A735 Exp: 10/31/2018 UDI: (01)04987595301636(17)181031(10)A735 and Lot: O736 Exp: 01/03/2019 UDI: (01)04987595301636(17)190131(10)O736
Based on the official FDA remedy

What to do — step by step

  1. Check Cartridge for Pinhole

    Inspect the cartridge for a pinhole in the aluminum sheet over the reagent wells.

Where it went and when

Where it was distributed

Distributed in 8 states.

Alaska: not listedAKMaine: not listedMEVermont: not listedVTNew Hampshire: not listedNHWashington: not listedWAIdaho: yesIDMontana: not listedMTNorth Dakota: not listedNDMinnesota: yesMNIllinois: not listedILWisconsin: not listedWIMichigan: not listedMINew York: yesNYRhode Island: not listedRIMassachusetts: not listedMAOregon: not listedORNevada: not listedNVWyoming: not listedWYSouth Dakota: not listedSDIowa: not listedIAIndiana: not listedINOhio: not listedOHPennsylvania: not listedPANew Jersey: not listedNJConnecticut: not listedCTCalifornia: yesCAUtah: not listedUTColorado: not listedCONebraska: not listedNEMissouri: not listedMOKentucky: not listedKYWest Virginia: not listedWVVirginia: not listedVAMaryland: not listedMDDelaware: not listedDEArizona: not listedAZNew Mexico: not listedNMKansas: not listedKSArkansas: yesARTennessee: not listedTNNorth Carolina: not listedNCSouth Carolina: not listedSCDistrict of Columbia: not listedDCOklahoma: not listedOKLouisiana: yesLAMississippi: not listedMSAlabama: not listedALGeorgia: not listedGAHawaii: not listedHITexas: yesTXFlorida: yesFL

FDA distribution: “US Distribution to states of: AR, CA, FL, ID, LA, MN, NY and TX.” States read from FDA’s own wording; “not listed” means the notice does not name it.

States: Arkansas, California, Florida, Idaho, Louisiana, Minnesota, New York, Texas.

Recall timeline

FDA terminated this recall 12 months after the firm started it.

1. Firm started the recall: Apr 30, 201812. FDA classified it Class II: Sep 17, 201823. Published in FDA’s Enforcement Report: Sep 26, 201834. FDA terminated the recall: Apr 17, 2019420182019
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-3185-2018 (April 30, 2018) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-3185-2018 to see it in context.
  • Plain-language summary written by qwen3:4b on August 21, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA