Synthes (USA) Products LLC recalled the Synthes Torque Limiting Handle, Part Number 03.231.013, because the adapter may demonstrate a degradation of function due to ball bearings likely to rust prematurely, according to the FDA. The agency announced the voluntary Class II recall on May 4, 2018.
How to tell if you have it
- Brand: Synthes
- Product: Synthes Torque Limiting Handle
- Part Number: 03.231.013
- Lot Codes: H344006-02 through H344006-41, H344006-43, H344006-45 through H344006-49, H369323-03, H369323-04
- Distribution: USA (nationwide); states include AK, CA, FL, ID, IN, KS, KY, LA, MD, MI, MT, NC, NJ, NY, OH, OK, PA, SD, TN, TX, UT, WI, and WV
The hazard
The device is used in systems indicated for fracture fixation. The adapter may demonstrate a degradation of function because the final products were assembled with ball bearings of high carbon content that is likely to rust prematurely.
Context
GovInfoNow's records show Synthes (USA) Products LLC has 19 other recalls on record since July 21, 2017. Other recent recalls of similar products include Synthes 2.4V TAP CORTEX SCREW Recall: Incorrect Thread Hazard (Feb. 25, 2026), Synthes LOW PROF LCKNG SCREW: Incorrect Length Labeling on Packaging (Sept. 26, 2025), and Synthes Stardrive Screwdrivers May Cause Surgical Delay - Recall (April 25, 2024).