Skip to content
Lower risk Medical device recall · FDA · Published May 16, 2018

BD FACS Universal Loader (Model 662879) May Not Lock Properly During Use

BD Biosciences recalls BD FACS Universal Loader (Model 662879) due to potential failure to lock during use, which could lead to improper sample handling. Affected units include specific serial numbers.

🇺🇸 FDA recall #Z-2598-2018Medical deviceBD Biosciences ›
Take action

Contact the company

Recalling firmBecton, Dickinson and Company, BD Biosciences2350 Qume Dr · San Jose, CA 95131-1812
✎ GovInfoNow plain-language explainer

What happened, in plain English

BD Biosciences is recalling the BD FACS Universal Loader (Model 662879) because the solenoid plunger tip is too short, which may prevent it from locking securely during use. This could lead to improper sample handling if the loader does not stay locked. Read more at govinfonow.com/recall/71057-bd-facs-universal-loader-model-662879-may-not-lock-properly.

Source Summary by GovInfoNow · govinfonow.com/recall/71057-bd-facs-universal-loader-model-662879-may-not-lock-properly · Updated August 21, 2026 · FDA enforcement record: FDA #Z-2598-2018

Cite this page

“BD FACS Universal Loader (Model 662879) May Not Lock Properly During Use.” GovInfoNow, updated August 21, 2026. https://govinfonow.com/recall/71057-bd-facs-universal-loader-model-662879-may-not-lock-properly. Based on FDA recall #Z-2598-2018 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2598-2018%22&limit=1).

Why it’s dangerous

The solenoid plunger tip is too short, which means it may not engage properly with the Universal Loader enclosure. This could cause the loader to become unlocked during use, potentially leading to improper sample handling or spills.

Who’s affected

Users of the BD FACS Universal Loader for clinical laboratory applications are affected. Those handling samples in 30- or 40-tube racks are most at risk if the loader fails to lock securely.

How to tell if you have this one

Check for the model number 662879 on the label. Affected units have specific serial numbers: 2662879000001 through 2662879000012.

  • Brand
    BD Biosciences
  • Product
    BD FACS Universal Loader (Optional) to the BD FACSLyric Clinical System, Model 662879 (Universal Loader), Rx Only. The
  • Category
    Medical device
Based on the official FDA remedy

What to do — step by step

  1. Check Serial Number

    Look for the serial number on the product label to determine if it is affected (2662879000001 through 2662879000012).

Where it went and when

Where it was distributed

Distributed in 6 states.

Alaska: not listedAKMaine: not listedMEVermont: not listedVTNew Hampshire: not listedNHWashington: not listedWAIdaho: not listedIDMontana: not listedMTNorth Dakota: not listedNDMinnesota: not listedMNIllinois: not listedILWisconsin: not listedWIMichigan: not listedMINew York: not listedNYRhode Island: not listedRIMassachusetts: yesMAOregon: not listedORNevada: not listedNVWyoming: not listedWYSouth Dakota: yesSDIowa: not listedIAIndiana: not listedINOhio: not listedOHPennsylvania: not listedPANew Jersey: yesNJConnecticut: not listedCTCalifornia: not listedCAUtah: not listedUTColorado: not listedCONebraska: not listedNEMissouri: not listedMOKentucky: yesKYWest Virginia: not listedWVVirginia: not listedVAMaryland: not listedMDDelaware: not listedDEArizona: not listedAZNew Mexico: not listedNMKansas: yesKSArkansas: not listedARTennessee: not listedTNNorth Carolina: not listedNCSouth Carolina: not listedSCDistrict of Columbia: not listedDCOklahoma: not listedOKLouisiana: not listedLAMississippi: not listedMSAlabama: not listedALGeorgia: not listedGAHawaii: not listedHITexas: not listedTXFlorida: yesFL

FDA distribution: “US Distribution was made to FL, KS, KY, MA, NJ, and SD.” States read from FDA’s own wording; “not listed” means the notice does not name it.

States: Florida, Kansas, Kentucky, Massachusetts, New Jersey, South Dakota.

Recall timeline

FDA terminated this recall 3 months after the firm started it.

1. Firm started the recall: May 16, 201812. FDA classified it Class III: Aug 2, 201823. Published in FDA’s Enforcement Report: Aug 8, 201834. FDA terminated the recall: Aug 13, 201842018
  1. Firm started the recall
  2. FDA classified it Class III
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-2598-2018 (May 16, 2018) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-2598-2018 to see it in context.
  • Plain-language summary written by qwen3:4b on August 21, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA