The FDA announced on May 16, 2018, that Zimmer Biomet, Inc. voluntarily recalled various trauma guide wires due to potential sterile barrier failure. The agency issued recall number Z-2237-2018 for the Class II device.
How to tell if you have it
- Brand: Zimmer Biomet
- Product: G-PIN 2.8BL 3.2PL 98CM 2PK; 467260
- Intended use: Guide pin for insertion of implants (cannulated screws and/or intramedullary nails)
The hazard
The firm recalled the guide wires due to insufficient packaging design verification and an exceeding expected occurrence rate of complaints for sterile barrier failure. The breach has the potential to lead to a possible risk for infection.
Context
GovInfoNow's records show Zimmer Biomet, Inc. has 94 other recalls on record since June 5, 2017. Other recent recalls of similar products include Zimmer Sterile Tourniquet Hoses Recall: Sterility Breach Risk (June 9, 2026), Merit Medical Splashwire Hydrophilic Guide Wire Recall: Unsealed Pouches Risk (May 8, 2026), and Biomet Intramedullary Bone Saw Blade Recall: Sterile Packaging Risk (July 1, 2026).