Stop using now Medical device recall · FDA · Published May 24, 2018

Draeger Perseus A500 Anesthesia Machine Recall: Risk of Patient Hypoxia

Draeger Medical recalls Perseus A500 anesthesia machines due to potential risk of delivering 100% nitrous oxide, which could cause patient hypoxia and death. Affected units have specific model numbers and were sold for medical use.

🇺🇸 FDA recall #Z-2254-2018Medical deviceDraeger ›
Take action

Contact the company

Recalling firmDraeger Medical, Inc.3135 Quarry Rd · Telford, PA 18969-1042
✎ GovInfoNow plain-language explainer

Draeger Medical is recalling Perseus A500 anesthesia machines because a fault in the S-ORC module could allow the device to deliver 100% nitrous oxide to patients, risking hypoxia and death. This is a serious hazard that requires immediate action to prevent patient harm. Read more at govinfonow.com/recall/70966-draeger-perseus-a500-anesthesia-machine-recall-risk-of-patient-hypoxia.

Source Summary by GovInfoNow · govinfonow.com/recall/70966-draeger-perseus-a500-anesthesia-machine-recall-risk-of-patient-hypoxia · Updated August 21, 2026 · FDA enforcement record: FDA #Z-2254-2018

Cite this page

“Draeger Perseus A500 Anesthesia Machine Recall: Risk of Patient Hypoxia.” GovInfoNow, updated August 21, 2026. https://govinfonow.com/recall/70966-draeger-perseus-a500-anesthesia-machine-recall-risk-of-patient-hypoxia. Based on FDA recall #Z-2254-2018 (https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2254-2018%22&limit=1).

The FDA announced on May 24, 2018, that Draeger Medical, Inc. is recalling the Perseus A500 Anesthesia Machine due to a hazard where the device may dose a hypoxic mixture of nitrous oxide and oxygen. This fault in the S-ORC module can result in patient death.

How to tell if you have it

  • Brand: Draeger
  • Product: Perseus A500 Anesthesia Machine
  • Catalog number: MK06000
  • Lot codes: ASKM-0197, ASKN-0006 through ASKN-0055 (excluding listed gaps), ASLA-0009, ASLA-0010

The hazard

According to the FDA, the S-ORC module in the Perseus A500 Anesthesia Machine may fail to prevent setting a nitrous oxide flow that results in a hypoxic mixture being dosed to the patient. The record states that potential adverse outcomes include death of the patient.

Context

GovInfoNow's records show Draeger Medical, Inc. has had 16 other recalls on record since March 1, 2018. Other recent recalls of similar products include the Draeger Atlan A350 Anesthesia Workstation Recall (May 7, 2026) and the Draeger Vapor 3000 Anesthetic Vaporizer Recall (Nov. 24, 2025).

How to tell if you have this one

Check for the model number MK06000 on the Perseus A500 anesthesia machine. These units were sold for medical use in the U.S. and intended for anesthesia of adults, children, and neonates.

  • Brand
    Draeger
  • Category
    Medical device
FDA recall event · 5 products

All 5 products in this recall

FDA lists each product line separately under one recall event. Check your product description, lot or code numbers against each row. Every row shares the hazard and remedy on this page.

  • Classification
    Class II
  • Status
    Terminated
  • Distribution
    US Nationwide Distribution - Healthcare facilities in AK, AL, AZ, CA, CO, FL, GA, IA, IL, IN, MA, MD, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, WA, WI, and WV.
Recall numberProduct, codes & lotsQuantity
Z-2250-2018Fabius GS Premium Anesthesia Machine; Cat. no. 8607000 Product Usage: Inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series machines are equipped with a compact breathing system; providing fresh gas decoupling, PEEP, and pressure limitation. The following ventilation options are available: Volume Controlled Ventilation Pressure Controlled Ventilation Pressure Support (Optional) SIMV/PS (Optional) Manual Ventilation Spontaneous Breathing Fabius series anesthesia machines are equipped with an electrically driven and electronically controlled ventilator and monitors for airway pressure (P), volume (V), and inspiratory oxygen concentration (FiO2). As per IEC 60601-2-13 (Anesthetic Workstations and their Modules-Particular Requirements), additional monitoring of the concentrations of CO2 and anesthetic agent is required when the machine is in use. IEC 60601-2-13:2003 requires that a manual ventilation bag must be available for emergency use. Fresh gas enrichment is provided by the Dr¿ger Vapor anesthetic vaporizer. Fabius MRI can also be used in MRI scanner rooms with magnets 1.5 tesla and 3 tesla at a fringe field strength of 40 mtesla or less; ASKM-0102 ASKM-0220 ASKM-0103 ASKM-0221 ASKM-0124 ASKM-0222 ASKM-0125 ASKM-0223 ASKM-0129 ASKM-0224 ASKM-0132 ASKM-0229 ASKM-0133 ASKN-0061 ASKM-0134 ASKN-0090 ASKM-0135 ASLA-0035 ASKM-0136 ASLA-0046 ASKM-0137 ASLA-0047 ASKM-0138 ASLA-0048 ASKM-0139 ASLA-0049 ASKM-0189 ASLA-0050 ASKM-0190 ASLA-0053 ASKM-0191 ASLA-0070 ASKM-0192 ASLA-0071 ASKM-0193 ASLA-0072 ASKM-0194 ASLA-0073 ASKM-0195 ASLA-0074 ASKM-0198 ASLA-0075 ASKM-0199 ASLA-0129 ASKM-0200 ASLA-0131 ASKM-0201 ASLA-0133 ASKM-0202 ASLA-0140 ASKM-0204 ASLA-0141 ASKM-0205 ASLA-0159 ASKM-0207 ASLA-0199 ASKM-0208 ASLA-0200 ASKM-0215 ASLA-0201 ASKM-0219 ASLA-021862
Z-2251-2018Fabius GS Tiro Anesthesia Machine; Cat. no. 8606000 Product Usage: Inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series machines are equipped with a compact breathing system; providing fresh gas decoupling, PEEP, and pressure limitation. The following ventilation options are available: Volume Controlled Ventilation Pressure Controlled Ventilation Pressure Support (Optional) SIMV/PS (Optional) Manual Ventilation Spontaneous Breathing Fabius series anesthesia machines are equipped with an electrically driven and electronically controlled ventilator and monitors for airway pressure (P), volume (V), and inspiratory oxygen concentration (FiO2). As per IEC 60601-2-13 (Anesthetic Workstations and their Modules-Particular Requirements), additional monitoring of the concentrations of CO2 and anesthetic agent is required when the machine is in use. IEC 60601-2-13:2003 requires that a manual ventilation bag must be available for emergency use. Fresh gas enrichment is provided by the Dr¿ger Vapor anesthetic vaporizer. Fabius MRI can also be used in MRI scanner rooms with magnets 1.5 tesla and 3 tesla at a fringe field strength of 40 mtesla or less; ASKM-0110 ASKM-0111 ASKM-0196 ASKM-0209 ASKM-0232 ASKM-0244 ASKM-0246 ASKN-0032 ASKN-0033 ASKN-0034 ASKN-0037 ASKN-0041 ASKN-0043 ASKN-0068 ASKN-0093 ASLA-0009 ASLA-0021 ASLA-0023 ASLA-0031 ASLA-0032 ASLA-0123 ASLA-0130 ASLA-0135 ASLA-0136 ASLA-0147 ASLA-0155 ASLA-0158 ASLA-0163 ASLA-0166 ASLA-0170 ASLA-019531
Z-2252-2018Fabius GS MRI Anesthesia Machine; Cat. no. 8607300 Product Usage: Inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series machines are equipped with a compact breathing system; providing fresh gas decoupling, PEEP, and pressure limitation. The following ventilation options are available: Volume Controlled Ventilation Pressure Controlled Ventilation Pressure Support (Optional) SIMV/PS (Optional) Manual Ventilation Spontaneous Breathing Fabius series anesthesia machines are equipped with an electrically driven and electronically controlled ventilator and monitors for airway pressure (P), volume (V), and inspiratory oxygen concentration (FiO2). As per IEC 60601-2-13 (Anesthetic Workstations and their Modules-Particular Requirements), additional monitoring of the concentrations of CO2 and anesthetic agent is required when the machine is in use. IEC 60601-2-13:2003 requires that a manual ventilation bag must be available for emergency use. Fresh gas enrichment is provided by the Dr¿ger Vapor anesthetic vaporizer. Fabius MRI can also be used in MRI scanner rooms with magnets 1.5 tesla and 3 tesla at a fringe field strength of 40 mtesla or less; ASLA-0017 ASLA-0018 ASLA-0019 ASLA-00204
Z-2253-2018Apollo Anesthesia Machine; Cat. no. 8605310 Product Usage: Indicated as a continuous flow anesthesia system. The Apollo may be used for manually assisted or automatic ventilation, delivery of gases and anesthetic vapor, and monitoring of oxygen and CO2 concentration, breathing pressure, respiratory volume, and anesthetic agent concentration and identification. It is an inhalation anesthesia machine for use in operating, induction, and recovery rooms. It can be used with rebreathing systems, semiclosed to virtually closed systems with low flow and minimal flow techniques, and non-rebreathing systems (with the Auxiliary Common Gas Outlet). It may be used with O2, N2O, and Air supplied by a medical gas pipeline system or by externally mounted gas cylinders. Anesthetic agent can be delivered via vaporizers mounted on the machine. The Apollo is equipped with a compact breathing system, providing fresh-gas decoupling, PEEP, and pressure limitation. It has an electrically driven and electronically controlled ventilator.; ASKM-0152 ASKN-0138 ASKM-0193 ASKN-0139 ASKM-0202 ASKN-0143 ASKM-0203 ASKN-0144 ASKM-0211 ASKN-0148 ASKM-0212 ASKN-0150 ASKM-0213 ASKN-0151 ASKM-0214 ASKN-0152 ASKM-0221 ASKN-0163 ASKM-0222 ASKN-0164 ASKM-0227 ASKN-0165 ASKM-0228 ASLA-0001 ASKM-0246 ASLA-0002 ASKM-0269 ASLA-0003 ASKM-0272 ASLA-0092 ASKM-0328 ASLA-0121 ASKM-0332 ASLA-0124 ASKM-0333 ASLA-0131 ASKM-0337 ASLA-0132 ASKM-0339 ASLA-0145 ASKM-0340 ASLA-0146 ASKM-0360 ASLA-0152 ASKM-0362 ASLA-0153 ASKM-0366 ASLA-0154 ASKM-0370 ASLA-0163 ASKM-0371 ASLA-0164 ASKM-0408 ASLA-0165 ASKM-0409 ASLA-0166 ASKM-0415 ASLA-0173 ASKN-0012 ASLA-0174 ASKN-0035 ASLA-0175 ASKN-0039 ASLA-0179 ASKN-0040 ASLA-0180 ASKN-0113 ASLA-0181 ASKN-0114 ASLA-0182 ASKN-0116 ASLA-018372
Z-2254-2018Perseus A500 Anesthesia Machine; Cat. no. MK06000 Product V Product Usage: Intended for use in anesthetizing adults, children, and neonates and can be used for automatic and manual ventilation, pressuresupported spontaneous breathing, and spontaneous breathing. Perseus is equipped with airway monitoring, gas measurement and device monitoring, O2 insufflation, and an anesthetic gas receiving system. Anesthesia is achieved through a mixture of pure oxygen and Air (medical compressed air) or pure oxygen and nitrous oxide, with the addition of volatile anesthetic agents. Ventilation is accomplished on the patient through a laryngeal mask airway, a full-face mask, or an endotracheal tube. The integrated breathing system can be used with partial rebreathing (low-flow or minimum-flow). A non-rebreathing system such as the Kuhn or Medec Water System may be used at the external fresh-gas outlet (optional). Perseus is specified for inhalational anesthesia and/or patient ventilation in accordance with the intended use during surgical or diagnostic interventions.; ASKM-0197 ASKN-0006 ASKN-0007 ASKN-0008 ASKN-0009 ASKN-0013 ASKN-0014 ASKN-0015 ASKN-0021 ASKN-0026 ASKN-0027 ASKN-0028 ASKN-0029 ASKN-0030 ASKN-0031 ASKN-0034 ASKN-0035 ASKN-0036 ASKN-0037 ASKN-0038 ASKN-0039 ASKN-0040 ASKN-0042 ASKN-0045 ASKN-0046 ASKN-0048 ASKN-0049 ASKN-0050 ASKN-0053 ASKN-0054 ASKN-0055 ASLA-0009 ASLA-001033
Based on the official FDA remedy

What to do — step by step

  1. Stop Using Immediately

    Discontinue use of the Perseus A500 anesthesia machine if it has the model number MK06000 to prevent potential patient hypoxia.

Where it went and when

Where it was distributed

Distributed nationwide; the notice names 33 states.

Alaska: named in the noticeAKMaine: nationwideMEVermont: nationwideVTNew Hampshire: named in the noticeNHWashington: named in the noticeWAIdaho: nationwideIDMontana: nationwideMTNorth Dakota: named in the noticeNDMinnesota: named in the noticeMNIllinois: named in the noticeILWisconsin: named in the noticeWIMichigan: nationwideMINew York: named in the noticeNYRhode Island: nationwideRIMassachusetts: named in the noticeMAOregon: named in the noticeORNevada: named in the noticeNVWyoming: nationwideWYSouth Dakota: nationwideSDIowa: named in the noticeIAIndiana: named in the noticeINOhio: named in the noticeOHPennsylvania: named in the noticePANew Jersey: named in the noticeNJConnecticut: nationwideCTCalifornia: named in the noticeCAUtah: named in the noticeUTColorado: named in the noticeCONebraska: named in the noticeNEMissouri: named in the noticeMOKentucky: nationwideKYWest Virginia: named in the noticeWVVirginia: nationwideVAMaryland: named in the noticeMDDelaware: nationwideDEArizona: named in the noticeAZNew Mexico: named in the noticeNMKansas: nationwideKSArkansas: nationwideARTennessee: named in the noticeTNNorth Carolina: named in the noticeNCSouth Carolina: nationwideSCDistrict of Columbia: nationwideDCOklahoma: named in the noticeOKLouisiana: nationwideLAMississippi: named in the noticeMSAlabama: named in the noticeALGeorgia: named in the noticeGAHawaii: nationwideHITexas: named in the noticeTXFlorida: named in the noticeFL

FDA distribution: “US Nationwide Distribution - Healthcare facilities in AK, AL, AZ, CA, CO, FL, GA, IA, IL, IN, MA, MD, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, WA, WI, and WV.” States read from FDA’s own wording; “not listed” means the notice does not name it.

Named in the notice: Alabama, Alaska, Arizona, California, Colorado, Florida, Georgia, Illinois, Indiana, Iowa, Maryland, Massachusetts, Minnesota, Mississippi, Missouri, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Tennessee, Texas, Utah, Washington, West Virginia, Wisconsin.

Recall timeline

FDA terminated this recall 2 years after the firm started it.

1. Firm started the recall: May 24, 201812. FDA classified it Class II: Jun 19, 201823. Published in FDA’s Enforcement Report: Jun 27, 201834. FDA terminated the recall: May 12, 2020420182020
  1. Firm started the recall
  2. FDA classified it Class II
  3. Published in FDA’s Enforcement Report
  4. FDA terminated the recall

Dates from the agency record only; steps without a date are left out.

Similar recalls

Latest medical device recalls

Keep exploring

Where this comes from

The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.

  • Official source: FDA Recall #Z-2254-2018 (May 24, 2018) — FDA publishes no per-recall page; this is the official machine-readable enforcement record. Search FDA recalls for Z-2254-2018 to see it in context.
  • Plain-language summary written by qwen3:4b on August 21, 2026, from the official record above. Not reviewed by a person — if it conflicts with the FDA notice, the notice is correct. Report an error on this page.
  • GovInfoNow is an independent service and is not a government agency. We republish official FDA recall data; we don’t issue, amend or administer recalls.
GovInfoNow is an independent guide. We are not FDA or any government agency. Always confirm details with the official recall.Data: CPSC · FDA · NHTSA · USDA