The FDA announced on July 12, 2018, that Ortho Clinical Diagnostics Inc. is recalling VITROS Immunodiagnostic Products - Folate Reagent Pack due to potential biased results for patients taking biotin supplements. The recall affects 7,880 units distributed worldwide.
How to tell if you have it
- Brand: Ortho Clinical Diagnostics
- Product: VITROS Immunodiagnostic Products - Folate Reagent Pack(10758750009237) 1513266
- Model numbers: 2470, 2480, 2490, 2501, 2510, 2521
- Quantity: 7,880 units
The hazard
The firm determined that biased results may occur for specific VITROS Immunodiagnostic Products at biotin concentrations lower than indicated in the current Instructions For Use. Patients taking biotin supplements could potentially have biased sample results for affected assays.
Context
GovInfoNow's records show Ortho-Clinical Diagnostics, Inc. has 53 other recalls on record since July 10, 2017. Recent brand recalls include calibration failures affecting VITROS Chemistry Products LAC Slides and VITROS Ca Slides in July 2025, and inaccurate test results from VITROS OP Reagent Gen 15 in June 2025. Other recent recalls of similar products include Vivoo Test Strips Recall (Nov. 22, 2025), Vivoo Protein Test Recall (Nov. 22, 2025), and Vivoo pH Test Recall (Nov. 22, 2025).