The FDA announced on July 13, 2018, that Fresenius Medical Care Renal Therapies Group, LLC voluntarily recalled the 2008T Hemodialysis System with BIBAG. The agency stated the recall addresses a hazard where disinfectant could remain in the machine and be delivered to patients during treatment.
How to tell if you have it
- Brand: Fresenius Medical Care
- Product: 2008T HEMODIALYSIS SYSTEM W/BIBAG, Part Number 190766
- Applicable only with 2008T BlueStar Premium Upgrade Kit, Part Number 191138
- Units affected: 40
- Sold nationwide in CA, CT, MI, NC, and NY
The hazard
According to the FDA, if a power failure or interruption occurs during the cleaning/disinfection program with the auto-start feature enabled, the machine may enter the dialysis program before the mandatory rinse is performed. As a result, disinfecting chemical may not be completely removed from the machine before starting a treatment and disinfectant could exist in the dialysate delivered to the patient.
Context
GovInfoNow's records show Fresenius Medical Care Renal Therapies Group, LLC has 15 other recalls on record since 2017-09-29. Other recent recalls of similar products include Fresenius Medical Care 5008X HD/HDF with Twister Tubing Set (2026-08-10), Fresenius Medical Care CombiSet Hemodialysis Tubing Set Recall (2026-07-29), and Fresenius Medical Care GRANUFLO Dry Acid Concentrate Recall (2026-07-14). Additional similar product recalls include Fresenius Medical Care 5008X CAREsystem Recall (2025-10-25), Fresenius 5008X Blood Tubing Set Recall (2026-05-11), and Vantive US Healthcare OXIRIS Dialyzer Recall (2026-01-06).