Tobii Dynavox recalls TD Navio Midi speech devices with model 1000798 due to potential battery short circuits causing heating and gas venting. Affected units have specific serial numbers.
Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.
Tobii Dynavox is recalling TD Navio Midi speech devices with model number 1000798 because battery mounting tabs can become damaged, leading to short circuits that might cause heating and gas venting. This could pose a fire risk if the device is used in a way that allows the battery to move freely.
The battery mounting tabs can become damaged under certain conditions, allowing the battery to move inside the device. This movement may cause an internal short circuit, leading to localized heating and gas venting within the device enclosure.
People who own or use the TD Navio Midi speech device with model number 1000798 and serial numbers from TDNM1-240628000004 to TDNM1-260617070001 are affected. Those who use the device for communication assistance are most at risk due to potential device failure.
Check for the model number 1000798 on the device. Affected units have serial numbers starting with TDNM1-240628000004 and ending with TDNM1-260617070001.
Look for the serial number on the device, which should start with TDNM1-240628000004 and end with TDNM1-260617070001.
Reach out to Tobii Dynavox for further instructions on how to handle the device.
The battery mounting tabs can become damaged, allowing the battery to move inside the device, which may cause an internal short circuit leading to heating and gas venting.
Model number 1000798 with serial numbers from TDNM1-240628000004 to TDNM1-260617070001.
The recall includes corrected/future production units with new serialization starting from TDNM1-260617070001 onward, but the record does not specify a repair or replacement process.
We’ve drafted a message you can send Tobii Dynavox to request your refund or repair — edit it as you like.
Subject: Recall Z-2959-2026 — Tobii Dynavox Tobii Dynavox Hello, I own a Tobii Dynavox that is covered by recall Z-2959-2026 from Tobii Dynavox. I’m writing to arrange the remedy described in the recall (a refund, repair, or replacement as applicable). Could you please tell me the next steps to resolve this? Thank you.
Experiences, tips, and questions from readers. Every comment is reviewed before it appears.
No comments yet — be the first to share your experience.
The facts (hazard, remedy, dates, manufacturer) are pulled from the official FDA recall record. The plain-language summary and step-by-step checklist are written by GovInfoNow to make the official notice easier to act on — they don’t replace it.