The FDA announced on 2018-09-26 that Oscor Inc. is recalling TB Series Bipolar Pacing Leads due to a hazard where connector caps may slide and expose wires, potentially causing breakage. The agency issued recall number Z-0391-2019 for the Class I device.
How to tell if you have it
- Brand: Oscor
- Product: TB Series Bipolar Pacing Leads
- Sizes: French Size 4F, 5F, 6F
- Curve Types: Straight, Atrial J, 60 degree Curve, Right Heart
- Connectors: 2mm unshrouded
- Model Numbers: 020004, 020005, 020006, 020010, 020011, 020012, 020016, 020017, 020018, 020022, 020023, 020024
- Lot Codes: See detailed list in FACTS (e.g., 2B-81-532Z-D, 2B-81-532Z-E, etc.)
- Units Affected: 36,688
The hazard
During use of some TB Temporary Bipolar Pacing Leads with 2mm unshrouded connectors, the connector cap housing may slide and potentially expose the connection wire. This exposure can make the wire susceptible to loss of connectivity or breakage during cable movement, causing interruption of the pacing system.
Context
GovInfoNow's records show Oscor Inc. has issued 3 other recalls since 2021-03-10. On 2026-06-10, Oscor Abiomed recalled an Introducer Kit due to potential sheath leakage risk. Other recent recalls of similar products include Olymp BiCOAG Hemostasis Probe Connector Recall (2026-08-19) and Medline NAMIC Angiographic Control Syringes Recall (2026-02-27).