Tosoh Bioscience Inc recalled AIA-2000 Analyzers (Model Nos. AIA-2000LA, AIA-2000ST) due to a potential liquid suction pump failure causing insufficient wash solution removal, according to the FDA on 2018-09-28.
How to tell if you have it
- Brand: Tosoh Bioscience
- Product: AIA-2000 Analyzer
- Model Nos.: AIA-2000LA, AIA-2000ST
- Serial Nos.: 10038409, 10038509, 10048009R, 10058011, 10069507, 80018112, 80039511
- Quantity: 7 units
- Distribution: US Nationwide; Florida, Kentucky, Maryland, Ohio, Pennsylvania, Texas
The hazard
The liquid suction pump may not perform as intended, causing an insufficient volume of wash solution to be removed during the washing step. This incomplete washing may result in falsely elevated test results for sandwich assays and falsely decreased results for competitive binding assays. There have been no user or patient injuries reported related to this issue.
Context
GovInfoNow's records show Tosoh Bioscience Inc has 19 other recalls on record since 2017-06-11. Other recent recalls of similar products include Tosoh ST AIA-Pack Testosterone Recall (2026-06-12), Tosoh AIA-Pack BHCG Calibrator Recall (2020-08-07), and Tosoh AIA-900 Analyzer Recall (2020-05-22). Other recent recalls of similar products include Siemens Atellica CH Urine Albumin Recall (2026-07-06) and Vivoo Test Strips Recall (2025-11-22).