The FDA announced on Nov. 16, 2018, that LivaNova USA Inc. is recalling The LivaNova VNS Therapy System due to inaccurate lead impedance readings. Owners are advised to contact the company for remedy details.
How to tell if you have it
- Brand: LivaNova
- Product: The LivaNova VNS Therapy System
- Model: Model 1000 Generator
- Serial Numbers: Less than 100000
The hazard
The affected VNS generators report higher lead impedance values than previous models because of a change in measurement timing during diagnostic testing. This shift causes normal impedance ranges to fall outside the existing labeling thresholds of 600 - 5300 Ohms.
Context
GovInfoNow's records show LivaNova USA, Inc. has 18 other recalls on record since Jan. 24, 2018. Other recent recalls include a Jan. 3, 2025 recall for implanted pulse generators potentially stopping therapy and a Dec. 12, 2024 recall for SenTiva VNS Generators that may stop working.