The FDA announced on 2018-11-16 that Invacare Corporation is recalling Invacare 9805 and 9805P Hydraulic Patient Lifts due to the potential for mounting bolts to loosen. This defect can cause the lift to tilt and a patient to fall, resulting in injuries ranging from bruising to death. Owners should contact Invacare for a remedy.
How to tell if you have it
- Brand: Invacare
- Product: 9805 and 9805P Hydraulic Patient Lifts
- Model Nos.: 9805 and 9805P
- Serial numbers: 09FSZ590001 to 16ESZ104075, including starting values 10ASZ through 16ASZ, 10BSZ through 16BSZ, 10CSZ through 16CSZ, 10DSZ through 16DSZ, 10ESSZ through 16ESSZ, 09FSZ through 15FSZ, 09GSZ through 15GSZ, 09HSZ through 15HSZ, 09ISZ through 15ISZ, 09JSZ through 15JSZ, 09KSZ through 15KSZ, and 09LSZ through 15LSZ
The hazard
Loose mounting bolts connecting the caster to the base frame can cause wear and eventual separation of the caster, leading the lift to tilt and a patient to fall. Injuries from such falls typically include bruising, lacerations, or abrasions, but can rarely result in bone fracture, head trauma, or death. Caregivers attempting to stabilize an unstable lift may suffer sprains, serious back injuries, or bone fractures.
Context
GovInfoNow's records show Invacare Corporation has had 6 other recalls since 2020-10-30. Recent Invacare recalls include beds with potential weld defects announced on 2024-01-22, DOLOMITE GLOSS Rollators with seat collapse risks on 2023-06-15, and Platinum 5NXG Oxygen Concentrators with missing labels on 2022-06-10. Other recent recalls of similar products include Arjo Tenor Patient Lift Recall (2026-03-10), Mirion Captus 4000e Thyroid Uptake System Recall (2026-07-20), and Molift Mover 205 Recall (2025-11-17).